Содержание Ti-Max X205L

Страница 1: ......

Страница 2: ...to the device Safety instructions are classified as follows in accordance with the seriousness of the risk Class W A R N I N G C A U T I O N N O T I C E Degree of Risk Hazard that could result in ser...

Страница 3: ...l Do not allow any impact on to the product Do not drop the product Operators and all others in the area must wear eye protection and a mask when operating this motor Should the product function abnor...

Страница 4: ...ectricity supply of the power source is below 15W both under normal and single failure conditions 2 The power source uses a SELV circuit for electricity supply 3 The output voltage of the power source...

Страница 5: ...Supply Pressure Measure the supply pressure at the handpiece hose connection point and set the pressure to the value specified on the specification table Fig 1 For Multi Gauge information refer to Op...

Страница 6: ...the motor Fig 2 When connecting an optic handpiece twist the handpiece until it locks into position with a clicking sound 2 Make sure the handpiece is firmly connected to the motor 4 2 Disconnection...

Страница 7: ...on Turn the FWD REV Change Ring to R Fig 4 3 When the FWD REV Change Ring is positioned at the middle of F and R the drive air is interrupted and the motor does not rotate Fig 5 5 Check before treatme...

Страница 8: ...e product never use any solvent such as benzine or thinner 7 Maintenance After each patient maintain the product as follows 7 1 Cleaning 1 Remove dirt and debris from the product Do not use a wire bru...

Страница 9: ...spray can upright 7 2 Lubrication NSK PANA SPRAY Plus Apply NSK PANA SPRAY Plus every time after each use and or before autoclaving 1 Remove the motor from the hose 2 Mount the arrow head spray nozzle...

Страница 10: ...h Seal the pouch 2 Autoclavable under the conditions below Autoclave for more than 20 min at 121 C or 15 min at 132 C or 3 min at 134 C 3 The motor should remain in the autoclave pouch until required...

Страница 11: ...humidity ventilation and sunlight The air should be free from dust salt and sulphur Do not heat or cool the product too quickly Rapid change in temperature could cause damage to the product If the st...

Страница 12: ...rings make sure they are inserted in the correct grooves as shown in the Fig 8 8 Replacing the O rings Replace the O rings if water is present in the exhaust air line This is an indication of possible...

Страница 13: ...the motor and check for abnormalities such as abnormal rotation vibration noise and overheating 10 Specifications X205L ISO 9168 Type3 E Type ISO 3964 25 000min 1 2 0N cm 0 20 0 25MPa 2 0 2 5kgf cm2...

Страница 14: ...ondensation Humidity 30 75 Atmospheric Pressure 700 1 060hPa Temperature 10 50 C Humidity 10 85 Atmospheric Pressure 500 1 060hPa Coolant Water Supply Coolant Air Supply Water Spray Type Optic Voltage...

Страница 15: ...has been tampered with by unqualified personnel or has had non NSK parts installed Replacement parts are available for seven years beyond discontinuation of the model This product is Autoclavable up...

Страница 16: ...l as the risks of environmental contamination caused thereof a surgeon or a dentist is required to confirm the equipment is sterile Ask specialist firms who are licensed to dispose of specially contro...

Страница 17: ......