Handpiece
Positioning Mark
Motor
Fig.1
Fig. 2
Black Dots
Align the Black Dots
Open
Close
Bur Lock Ring
CAUTION
• When operating the handpiece always consider the safety of the patient.
• This handpiece is designed for implant surgery.
• Do not disassemble or alter the handpiece.
• Check for vibration, noise and overheating outside the patient's oral cavity before use. If any
abnormalities are found, stop using the handpiece immediately and contact authorized dealer.
• Should the handpiece function abnormally, cease operation immediately and return the handpiece to an
authorized dealer for repair.
• Do not allow any impact on to the handpiece. Do not drop the handpiece.
• Do not rotate the motor when the Bur Lock Ring is at OPEN position, or a bur is not mounted in the
chuck and do not turn the Bur Lock Ring while the handpiece is rotating. It may cause the
motor/handpiece disconnection or sudden heat generation.
• Do not connect or disconnect the handpiece until the derive motor has completely stopped.
• After each use, be sure to clean, lubricate and sterilize the handpiece. If blood is left on the handpiece, it
can clot causing rust.
• Perform periodical maintenance checks.
• The user shall be responsible for operation, maintenance and operation.
CAUTION for Bur Usage
• Only use suitable burs.
• Do not use bent, damaged or sub-standard burs. The Shank could bend or brake even within the
recommended speed.
• Always keep the bur shanks clean. Dirt or debris in a chuck could cause poor bur concentricity or
low chuck retention force.
• Follow the instructions of bur manufactures for recommended operating speed as bur breakage may
occur if rotated at an excessive speed.
• Always mount the bur even when the handpiece is not in use.
• Before use always check that the bur is correctly seated and locked in place.
• Make sure if the Black Dots of the Bur Lock Ring align after fixing the bur. Rotating the bur in an
improper position may cause heat generation, which results in burns.
3. Mounting and Removing Bur
(1) Mounting Bur
1) Turn the Bur Lock Ring in the direction of mark “O” until it
clicks.
2) Insert the Bur
3) Turn the Bur Lock Ring in the direction of mark “C” until it
clicks. Make sure the Black Dots align. (See the picture
below)
(2) Removing Bur
Turn the Bur Lock Ring in the direction of mark “O” until it
clicks and remove the bur.
2. Connecting / Disconnecting the Handpiece
(1) Connecting
Align the Positioning Marks of the handpiece and the motor. Insert those straight until end of the motor
insertion.
(2) Disconnecting
Hold the handpiece and the motor and pull those out straight.
1. Specifications
External Coolant Straight Handpiece Titanium Body
OM-H0133E
001
Ti-SG65L
Model
Gear Ratio
Allowable Max. Speed
Bur Size
Optics
Weight
Ti-SG65L
1 : 1 Direct Drive
40,000 min
-1
Handpiece Bur (ø2.35)
Glass Rod Optic
66g