SURGERY SYSTEM
OPERATION MANUAL
OM-E0262E
0 1 9 7
Страница 1: ...SURGERY SYSTEM OPERATION MANUAL OM E0262E 0197 ...
Страница 2: ...ion tube as medical waste after use The system functions normally in the environment where the temperature is at 0 40ºC 32 104ºF humidity at 10 85 RH atmospheric pressure at 500 1060hPa and no moisture condensation in the Control Unit Use at outside of these limits may cause malfunction S Classification of equipment Type of protection against electric shock Class l equipment Degree of protection a...
Страница 3: ...is key is to select match ratio of the handpiece attachment before use to the unit Press this key until the LCD display exhibits the correct gear ratio of the handpiece attachment h Coolant Flow key This key is used to select the coolant solution flow volume 5 flow volume rates are available for selection plus the flow can be turned off j Forward Reverse key This key is used to change the rotation...
Страница 4: ...e relevant to the preset maximum speed or torque When all bars illuminate the operating speed or torque is at maximum When bars are half illuminated then the operating speed is approximately 50 of the preset speed S LCD display on the unit console r t CAUTION The LCD display panel is produced from liquid crystal and should always be treated with care a Coolant Solution Flow Volume button This butt...
Страница 5: ...1 A click is heard when the motor cord plug is correctly inserted into the control unit To disconnect the plug pull back the lock joint then disconnect the cord Fig 2 5 2 Connecting the foot control Face the screw on the foot pedal control cord plug downward then insert the plug into the Foot Control cord jack on the Control Unit Secure the plug by fastening the lock nut See Figs 3 and 4 5 3 Conne...
Страница 6: ...s its upper or lower limit When the torque setting reaches the upper or the lower limit an audible beep is heard and the torque cannot be changed any further h Select the rate of coolant solution flow volume Select the rate of the coolant solution flow volume by pressing the Coolant Flow key The rate of coolant solution flow volume has 5 flow rate steps plus no coolant flow j Memorize settings Aft...
Страница 7: ...operation of the micromotor when the drilling load reaches the programmed torque upper limit the integrated torque limiter automatically activates in to prevent torque application excessive to the set requirement When the torque limiter activates the motor stops over beeping after 5 seconds To reactivate the micromotor simply press the System key h Stopping the micromotor Release the foot control ...
Страница 8: ...ined wash off with water and dry with a soft cloth Do not immerse the entire handpiece in water and prevent water from entering from back end of the handpiece Using a Spray Lubricant As shown in Fig 19 insert the spray lubricant can E Type nozzle into the back of the handpiece Spray the lubricant into the handpiece 2 3 times for 2 3 seconds each time If the handpiece head is excessively stained wi...
Страница 9: ...ut Voltage Dimensions Weight 200 40 000min 1 rpm DC30V ø 24 x L120mm 133g Without the motor cord NE111 AC120 230V 50 60Hz 48VA 75mL min 02 gal min W268 x D230 x H103mm 3 1kg C293 Y200 830 Y200 890 Y200 880 Y200 840 Y200 850 C221 C332 C295 H084 H083 SGM I SGM E16RI SGM E20RI SGM E32RI SGM E64RI SGM E256RI FBG E ES 85 ES 95 SGS E SGA E Min 12 5 10 6 25 3 125 1 56 800 1 000 1 000 200 200 Max 2 500 2 ...