OM-T0412E 002
High Speed Air Turbine Handpiece
Caution
· When operating the handpiece always consider the safety of the patient.
· Do not exceed maximum bur length.
· Check for vibration, noise and overheating outside the patient's oral cavity before use. If any abnormalities
are found, stop using the handpiece immediately and contact dealer.
· Should the handpiece function abnormally, cease operation immediately and return the handpiece to
dealer for repair.
· Depressing the Push-Button while the handpiece is rotating will result in OVERHEATING of the handpiece
head. Special caution must be exercised during use to keep cheek tissue AWAY from the Push-Button of
the handpiece. Contact with cheek tissue may cause the Push-Button to depress and burn injury to the
patient may occur.
· Do not exceed the Air Pressure recommended specification below as bur breakage or wear may occur if
it exceeds the Air Pressure.
· Do not allow any impact on to the handpiece . Do not drop the handpiece.
· Refer to the coupling manual before connecting it to the handpiece tubing.
· Do not attempt to disassemble the handpiece or tamper with the mechanism.
· NSK has not recommended the repair and disassembly of the cartridge without exception. Handpiece
may perform abnormally (abnormal noise or abnormal vibration). Damage, failure or accidents are outside
of our guarantee.
· Do not fixate the eyes of LED light. (pico BLED, pico WLED)
· If any abnormalities (dark, do not light on or flashing) are found for LED, stop using the handpiece
immediately and contact dealer. (pico BLED, pico WLED)
· This handpiece is classified as Class 1 LED Product. (pico BLED, pico WLED)
· Use a power source which meets the following requirements. (pico BLED, pico WLED)
1. The electricity supply of the power source is below 15W both under normal and single failure
conditions.
2. The power source uses a SELV circuit for electricity supply.
3. The output voltage of the power source is within the range recommended by the manufacturer of this
product.
Caution
Use only clean water to maintain the Clean Head System.
Caution
· Fully depress the Push Button and insert the bur into the chuck until it is secure.
· Do not use carbide bur when the crown is removed. It may cause malfunction.
· Do not apply too much pressure to the bur. Bur may break or bent or difficult to remove. (Be sure to
irrigate water. Cutting should be soft touch.)
· Do not use any bur longer that the NSK recommended Specification.
· Do not use bent, worn, damaged, or non-concentric burs. Such burs can cause damage to the
handpiece.
· Do not exceed the bur speed recommended by the bur manufacturer.
· Always keep the bur shank clean. Entry of hard debris into the chuck via the bur shank could cause
rotation slip and also prevent the bur from being securely located in the chuck.
2. Insertion & Removal of the bur
2-1 To insert the Bur
1) Insert the bur until it stops.( )
2) Depress the Push Button.( )
3) Insert the bur into the chuck until it stops.( )
4) Release the button.
2-2 To remove the Bur
Depress the Push Button and remove the bur.( )
Within this Operation Manual, Model will be written down as follows.
S-Max pico = pico, S-Max pico KL = pico KL, S-Max pico SL = pico SL, S-Max pico BLED = pico BLED,
S-Max pico WLED = pico WLED, S-Max pico WHL = pico WHL, S-Max pico MWL = pico MWL, S-Max pico STL = pico STL
KaVo
®
and MULTIflex
®
are registered trademarks of Kaltenbach & Voigt GmbH & Co., Germany. Sirona
®
is a registered trademark of Sirona Dental Systems GmbH, Germany.
W&H
®
and Roto Quick
®
are registered trademarks of W&H Dentalwerk Bürmoos GmbH, Austria. Bien-Air
®
and Unifix
®
are registered trademarks of Bien-Air Dental S.A.,Switzerland.
1. Specification
Intended to Use:
This handpiece is designed only for dental treatment use.
3. Maintenances
3-1 Cleaning (Clean-Head System)
1) Brush off the debris at the Clean Head Holes. Do not use a
wire brush. (Fig. 2)
2) Half fill a cup with clean water.
3) Rotate the handpiece and immerse half of the handpiece
head. (Fig. 3)
4) Rotate and stop intermittently the handpiece about 2 to 3
seconds. (About 3 times)
5) By the function of the clean head system, the clean head
hole can be washed.
6) Stop the handpiece and wipe it dry. If the dirt could not
remove from the hole, remove the cartridge and clean it by
brush.
7) Perform lubrication and sterilization. (3-6 Lubrication and 3-7
Sterilization)
3-2 Cleaning (Spray Port)
When spray nozzle are clogged, or spray does not exit evenly from three ports, clean the ports as
follows:
1) Remove the bur from the chuck.
2) Take out the cleaning wire from the back end of the brush
holder. Insert the wire straight into the spray ports with
caution.
Do not forcibly insert the wire into the port.
Damaged or disoriented port could cause the spray diverge
or directed away from the bur, and the cooling efficiency
deteriorates.
3) After use, clean the cleaning wire. Push into the brush holder
with the pointed end of the wire inward.
3-3 Cleaning (Handpiece)
This handpiece can be washed via Thermo Disinfector.
Scrub dirt and debris from the handpiece, and wipe clean with alcohol-immersed cotton swab or cloth.
Do not use a wire brush. Refer the manual of Thermo-Disinfector for performing it.
Caution
· Be sure to hold the handpiece firmly to prevent the handpiece from slipping out of hand by the spray
pressure when lubricating.
· Supply lubricant until it comes out of the handpiece head.
· Hold the PANA SPRAY Plus/PANA SPRAY can upright.
· Repeat lubrication until handpiece is clean and free of debris. If blood is left on the handpiece, it can
clot causing rust. This may cause failures those are shortage of rotation or torque.
Caution
· Do not autoclave the handpiece, even when it is bagged, together with other instruments, from
which chemicals are not sufficiently cleaned. This is to avoid discoloration and/or damage to the
plating of the handpiece.
· Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
Caution
· Make sure that the O-Ring is removed, when removing the Cartridge. O-Ring might be remained in
the Head, when removing the Cartridge. If there is a remained O-Ring, the new Cartridge can not be
inserted properly.
· Use only NSK genuine cartridge. If you use other cartridge, specification may not be satisfied and we
can not warranty of this situation.
· The head cap screw thread is very fine. To prevent damage to the thread always first tighten the
handpiece head cap using fingers, then secure with the head cap wrench.
· Make sure that the O-ring (Fig. 8) is correctly located on the cartridge.
Caution
· Do not disassemble the Taper Ring except the O-ring changing.
· Make certain that the Taper Ring is firmly tightened. If the Taper Ring is loose water and air leakage
could occur.
Caution
Make certain that the Taper Ring is firmly tightened. If the Taper Ring is loose water and air leakage
could occur.
6. Non-Retraction Valve
A Non-Retraction Valve shuts off the water retraction right at the handpiece head, to prevent oral fluids
sucked into the water line.
NSK Coupling
Remove the back-end Gasket. Pull and
remove the water tube, and replace the
Non-Retraction Valve. (Fig. 11)
7. Disposing product
Consult with dealer from whom you purchased it about waste disposal.
pico BLED
1) Remove the handpiece from the coupling
and loosen the Taper Ring to remove.(Fig. 9)
2) Pull out the joint then remove the
Non-Retraction Valve. Pull out the
old Non-Retraction Valve with thin
stick.(Fig. 12)
3) Insert New Non-Retraction Valve
securely.
4) Mount the Joint into the
handpiece. Make certain the pipe
and joint together. (Fig. 13)
5) Tighten the Taper Ring tight.
5. Replacing the O-Rrings (pico)
Replace the O-Rings if water is present in the exhaust air line. This is an indication of possible water
leakage within the coupling. ALWAYS change the complete set of O-Rings.
1) Loosen and Remove the Taper Ring at the rear of the handpiece. (Fig. 9)
2) Gently remove each O-Rings with your thumb. (Fig. 10)
3) Insert the complete set of new O-Rings in the correct grooves.
4) Replace and firmly tighten the Taper Ring.
*PTL O-ring Set: Order Code Y900580
4. Replacing the Cartridge
1) Insert a dummy bur into the chuck.
2) Unscrew the Head Cap by turning
counter-clockwise and remove.
3) Push up the dummy bur upwards to
remove Cartridge.
4) Clean the interior of the Head casing.
5) Insert the new Cartridge.
6) Replace the Head Cap and firmly screw into
place.
*SX-PU03: Order Code P1140200
Caution
Do not use a sharp tool to clean the cellular glass optic rod. It
could damage the glass and reduce the light transmission. If
illumination becomes dim please contact dealer.
3-4 Cleaning (Optic)
Wipe clean the Glass Rod (Fig.5) with an alcohol-immersed
cotton swab. Remove all debris and oil.
3-5 Cleaning (Rear of the handpiece) (pico STL)
Wipe all internal surfaces at rear of the handpiece with an
alcohol-immersed cotton swab.
3-6 Lubrication
Supply PANA SPRAY Plus/PANA SPRAY or Care3 Plus after
each use and/or before autoclaving.
PANA SPRAY Plus
1) Attach Spray Nozzle onto the PANA SPRAY Plus/PANA
SPRAY.
2) Shake the spray can well to mix the oil and the propellant.
3) Insert the Spray Nozzle in rear of the handpiece and spray at
least 2 seconds.
3-7 Sterilization
Autoclave sterilization is recommended.
Sterilization is required first time you use and after each patient as noted below.
AUTOCLAVING
1) Wipe dirt and debris from the handpiece, and wipe clean with alcohol-immersed cotton swab or
cloth. Do not use a wire brush.
2) Lubricate the handpiece. (Refer 3-5 Lubrication)
3) Insert the handpiece in the autoclave pouch and seal it.
4) Autoclavable up to a max. 135˚C.
ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
5) Keep the handpiece in the autoclave pouch to keep it clean until you use it.
* Sterilization at 121˚C for more than 15 minutes is recommended by EN13060 or EN ISO17665-1.
NSK automatic handpiece cleaning and lubrication system
When using NSK automatic handpiece cleaning and lubrication system, refer to the system instructions.
Loosen
Taper Ring
Water Tube
Non-Retraction Valve
Gasket
Joint
Non-Retraction Valve
Cotton Swab
S-Max pico
1
2
3
2 4
Clean Head Hole
Clean Head Hole
Clean Head Hole
Cleaning Wire
Spray Ports
Cotton Swab
Glass Rod
PANA SPRAY Plus/PANA SPRAY
Spray Nozzle
Wrench
Cartridge
Head Cap
Head
O-Ring
PUSH
O-Ring
Handpiece
Joint
Align
Handpiece
2018.08.23 002 S
Model
Head Type
Rotation Speed
Spray Type
Chuck Type
Bur
Max. Bur Length
Bur Fixing Length
Maximum Bur Dimension
Air Supply
Air Consumption
Head Dimension x Height
Optic
Voltage
Activate Voltage
Consumption Current
Use Environment
Transportation and
Storage Environment
pico
0.20MPa (2.0 kgf/cm
2
)~
0.25MPa (2.5 kgf/cm
2
)
0.25MPa (2.5 kgf/cm
2
)~
0.30MPa (3.0 kgf/cm
2
)
0.20MPa (2.0 kgf/cm
2
)~
0.25MPa (2.5 kgf/cm
2
)
Miniature
380,000 ~ 450,000 min
-1
Single Spray
Push Button Chuck
ISO1797-1 Type3 Ø1.59~1.60mm Short Shank Bur/Super Short Shank Bur
18.5 mm
8.0 mm
Ø 2.0 mm
pico KL / pico SL
45 ± 5 NL/min
Glass Rod
-
-
-
White LED
AC/DC 3.3 ± 0.05 V
3.0 - 4.0 V
0.38 A Typical (3.3V)
Glass Rod
-
-
-
Temperature : 0 - 40 ˚C, Humidity : 30 - 75% HR, Atmospheric Pressure : 700 - 1,060hPa
Temperature : -10 - 50 ˚C, Humidity : 10 - 85% RH, Atmospheric Pressure : 500 - 1,060hPa
Ø 8.6 mm x 9.0 mm
40 ± 5 NL/min