PRECAUTIONS

CAUTION

Fig. 1

Fig. 2

Fig. 3

Fig. 4

• Patient's safety is the first priority during treatment.
• Phatelus Light Pack is to be used by qualified personnel for 

dental treatment only.

• If abnormal conditions arise during use, stop immediately and 

contact your dealer.

• Never disassemble or modify.
• Do not subject unit to strong shocks.
• Do not pull on the AC adaptor cord.
• Never cut or shorten the AC adaptor cord because voltage will 

increase and cause the lamp to fail easily.

1. SET-UP DIAGRAM

2. CONNECTION

1)  Assembling in Dental Unit

If AC/DC24V terminal is available inside of the dental unit, AC 
adaptor is not required.  Take out AC or DC24V power source line 
from the dental unit and insert the end of the line into the hole of 
Light Pack marked IN.  If the power source is DC24V, no 
consideration need be given to the plus and minus polarities.

2)  Connection of AC Adaptor Lead Wire

Connect the Lead Wire from the AC adaptor to the hole maked IN 
as shown in Fig.1.
Insert the Lead Wire from the AC adaptor into the hole located on 
the side of Light Pack.  Insert the screw driver provided in the hole 
marked IN to release the spring clip and push in the lead wires to 
connect.  (Fig.2)

This AC adaptor has the 
polarity of plus and minus 
but the lead wire can be 
connected to any hole in 
any order marked IN.

3)  Connection of Lead Wire from Tubing

Insert the Lead Wire from tubing into the hole marked A or B in the 
same manner as you connected the Lead Wire from AC adaptor.
When connecting two Lead Wires from two tubings, use both the 
A and B holes, one wire for each hole.

4)  Installation of T-joint

Cut the drive air tubing and install T-joint.  (Fig.3)
Connect by-pass tubing from T-joint to the same letter nipple as 
the lead wire.
If installing 2 handpieces, the letters of the lead wires and the by-
pass tubings must be matched.
Application of a small amount of oil on the nipple makes it easy to 
insert the by-pass tubing.

5)  Installation of Light Pack

Install the Light Pack out of the way, underneath the work table or 
on the underside of the arm with double faced tap.
Fasten the lead wires with the clips provided so they do not hang 
down.  (Fig.4)

Please read this Operation Manual carefully and file for future reference.

PHATELUS LIGHT PACK

OPERATION MANUAL

OM-E0039E Rev.2

Never cut the lead wire even it is too long because the 
voltage will increase and the lamp will fail easily.
If the wire is too long, please bundle the remaining wire 
and place where it will not be desturbed.

Tubing PTL-4H

AC Adaptor

AC Adaptor Lead Wire L=5,000

B Illuminance
 Adjustment Knob

T-joint

B Illumination Time Delay

 Adjustment Hole

Chip Air Tubing

Water Tubing

Drive Air Tubing

A Illuminance Adjustment Knob

A Illumination Time Delay
                    Adjustment Hole

Lead Wire

L=700

Clip

Clip