17

 

To unlock the door before opening it, press the key

 

Start/Stop

.  

The cycle is over and the load can be taken out .  

 

ATTENTION: instrument and chambers are hot 

 

If a label printer is used, at the end of the cycle the display will ask to select the number of labels need-
ed, select the value with UP and DOWN, than opening the door the printer will release the labels. 

Opening the door, the displays will  show  again  the  current  time, chamber  temperature  and  pressure, 
and the sterilizer is ready for a new cycle. 

If an external serial printer is connected, a report will be issued during the cycle phases with the more 
significant data; the report can be filed as proof of the sterilization process performed. 

 

The operator can arrange other load on the trays and start a new sterilization cycle. with the advantage 
of  shorter  heating-up  time  as  the  chamber  is  already  warm,  or  press  key 

Power

  to  put  the  unit  in 

stand-by status 

(OFF on the display)

 

If  the  door  is  not  opened  or  a  key  pressed  within  30  minutes,  the  unit  switches  to  stand-by 
(OFF) automatically

WARNING

Should any failure or error occur during the cycle, Led 

Alarm 

turns on, the display Time will show the 

type  of  alarm  (see  chapter  ALARM)  and  the  door  remains  locked.  To  unlock  the  door,  press  key 

Start/Stop

 

ATTENTION: instrument and chambers are very hot. 

Contamination risk. 

 

 

4.4 

STOPPING THE CYCLE 

To stop the sterilization cycle, press key 

Start/Stop

T

he  display Time shows the message “

MANUAL STOP

”. 

Before opening the door, make sure that the display Pressure is showing the value 0.  A safety device locks 
the handle when the chamber under pressure. To unlock the door press key 

Start/Stop

Remove the load and check for the presence of water into the chamber. In case of wrapped instruments, we 
suggest to replace with new bags. 

Before loading the chamber again, dry it carefully and wait 10 minutes to allow the water to evaporate and be 
drained completely. 

 

Содержание iClave Plus

Страница 1: ...Rev 1 Date June 2015 plus Water Steam Sterilizer Class B OPERATION MANUAL...

Страница 2: ...all the agent Thanks for the confidence given All rights reserved No portion of this publication can be printed transmitted rewritten stored in a data recovery system translated in any foreign or comp...

Страница 3: ...1 Available sterilization cycles 15 4 3 2 Starting a sterilization cycle 16 4 4 STOPPING THE CYCLE 17 4 5 TOPPING UP AND DRAINING THE WASTE WATER TANK 18 4 5 1 Topping up the main tank 18 4 5 2 Drain...

Страница 4: ...t damages or missing parts see delivery note IN CASE OF DAMAGES OR MISSING PARTS PLEASE IMMEDIATELY INFORM AND IN DETAIL THE FORWARDER NSK AND ITS AREA AGENT 1 2 CONFORMITY TO EUROPEAN DIRECTIVES The...

Страница 5: ...3 1 3 PACKAGE DIMENSIONS AND WEIGHT Total weight 58 Kg Store the package for possible future shipment 1 4 UNPACKAGING...

Страница 6: ...1 OVERALL DIMENSIONS 2 2 USABLE SPACE IN THE CHAMBER Diameter 240 mm Depth 384 mm Capacity 17 5 l Useful dimensions per tray 315 x 214 mm x 2 315 x 168 mm x 2 Useful tray volume 10 l Net weight 47 Kg...

Страница 7: ...heard pull it gently you have to feel that the cap is free to move The valve can not be adjusted or serviced for safety reasons Follow the maintenance program described in the ap pendix SERVICE BOOK...

Страница 8: ...ent sterilization cycle has not been successfully completed The water waste contained in the recovery tank is to be regarded as biologically contaminating therefore when this tank is emptied suitable...

Страница 9: ...fresh water tank drain Quick connector for waste water tank drain Water inlet filter Temperature sensor Air line filter Safety valve connection Safety valve Automatic draining of used water tank opti...

Страница 10: ...y may be replaced only by au thorized service Varta CR2032 Sterilization time from 3 to 90 depending on the selected cycle Transmitted heat in environment 0 22 J h at 23 C Drying time from 6 to 14 dep...

Страница 11: ...ater easier during the drain phase WARNING do not place the unit on surfaces which could cause fire or fume if hot items fall from the equipment 3 For the correct operation it is mandatory to leave a...

Страница 12: ...he trays from the chamber and close the door When the unit is off the door is blocked if the block per sists turn off and then on again the unit 4 Hold on the key UP and push the key Power the display...

Страница 13: ...the factory test the equipment is set at a default value of 100m and can be left unchanged for real alti tude values between 0 and 200m since a 100m error does not affect the equipment operation To b...

Страница 14: ...ake another cycle Phase in progress signalling Led turned on or flashing during the phases of the cycle LCD displays Visualize from the top the value of the parameters Time when the unit is OFF Temper...

Страница 15: ...e numbers to the right indicate the value of the temperature in the chamber expressed in Celsius degrees Cycle class indicators B or S Temperature cycle indicators 121 C for cycles 2 and 7 134 C for c...

Страница 16: ...a fata backup The size of the data is limited to few Kbyte so a small capacity is needed a big amount of data stored in the USB stick decreases the speed of access so it is suggested to delete the fil...

Страница 17: ...mperature reading on the display is inaccurate because no steam being there 3 Arrange the material to be sterilized on the trays load the chamber and close the door 4 Check the indication of the water...

Страница 18: ...values icon Vacuum turns off and Led Sterilize turns on The display time starts the countdown marking the time remaining to the end of the sterilization phase The sterilization phase is followed by th...

Страница 19: ...ready warm or press key Power to put the unit in stand by status OFF on the display If the door is not opened or a key pressed within 30 minutes the unit switches to stand by OFF automatically WARNING...

Страница 20: ...n again only af ter one cycle or if the unit is switched off then the icon will blink again with the empty symbol when there is not enough water to run another cycle the current cycle will be anyway c...

Страница 21: ...value than push START the display shows SPECIAL TIME using UP and DOWN select the value than push START the display shows VACUUM SPECIAL using UP and DOWN select the value than push START the display...

Страница 22: ...umber of prints needed with UP and DOWN the print starts opening the door LANGUAGE selects the language of the display messages and cycles report ALTITUDE adjust the elevation above the see level of t...

Страница 23: ...avoid excessive heating To close the door switch on the unit again 1 Put a cleaning tab in the chamber and close the door THE CHAMBER MUST BE EMPTY 2 Select the cycle C CLEANING 3 Push START this cyc...

Страница 24: ...s is EN13060 ch 10 6 for the B type cycles and ch 10 5 for the S type For any explanations or information contact your authorized reseller or directly NSK 6 4 1 Porous load test B D The test can be pe...

Страница 25: ...ture and with the main tank full At the end of the auto test the display will show the message H2O good or H2O hard if the meas ured conductivity value is respectively lower or higher than 15 S ATTENT...

Страница 26: ...the mes sage appears before starting a cycle Top up the main tank FULL H2O The waste water tank is full the message ap pears before starting a cycle Drain the waste water tank MANUAL STOP Cycle manua...

Страница 27: ...1 8 4 ABORTED CYCLE ALARMS The alarms shown on table C refer to faults keep the sterilization process from being completed Identify the fault on the table and carry out the recommended action The alar...

Страница 28: ...aulty Call for a technical service AL 15 Chamber temperature sensor faulty Call for a technical service AL 16 Pressure sensor faulty Call for a technical service 8 5 CLASS B ADDITIONAL ALARMS TABLE D...

Страница 29: ...le development and with the following data content date and time of the process progressive cycle number selected program and parameters cycle type sterilization or disinfecting start end time of the...

Страница 30: ...e technician is able to detect any significant data of the machine and to per form more quickly and cheaply both troubleshooting and repairing Moreover the system allows the connection via external mo...

Страница 31: ...Standard periodic checks Pos Check Activity Ordinary maintenance Special maintenance 1 Door adjustment 1 Year 1 500 cycles Gasket replacement 1 Year 1 500 cycles Disk door clearance check 1 Year 1 500...

Страница 32: ...es Aborted cycles Alarm code Technician name Door adjustment Filter Calibration Cond cleaning Pump Tank cleaning Notes Sterilizer OK YES NO Date No of cycle Cleaning cycles Aborted cycles Alarm code T...

Страница 33: ...es Aborted cycles Alarm code Technician name Door adjustment Filter Calibration Cond cleaning Pump Tank cleaning Notes Sterilizer OK YES NO Date No of cycle Cleaning cycles Aborted cycles Alarm code T...

Страница 34: ...les Aborted cycles Alarm code Technician name Door adjustment Filter Calibration Cond cleaning Pump Tank cleaning Notes Sterilizer OK YES NO Date No of cycle Cleaning cycles Aborted cycles Alarm code...

Страница 35: ......

Страница 36: ...ht to make any altera tions which may be due to technical im provements Sous r serve de modifications dues au progr s tech nique Reservados los derechos de modificati n en virtud del progreso t cnico...

Отзывы: