Rubidium Status
– Discipline State
For reference, the Rb oscillator derives its frequency tuning algorithm constants
from its reference to the GNSS PPS discipline. When the algorithm has sufficient
averaging time from the available PPS, it indicates “DGOOD” for discipline good.
At this point, the NR300 is ready to enter holdover mode.
Prior to discipline good, the Rubidium indicates “DISCP”, which indicates a valid
PPS signal is present, and tuning is in process.
When no valid PPS is present, the Rubidium indicates that the unit is in holdover
condition. The Rubidium considers holdover to be the default state any time a
PPS reference is absent.