Operation
The NR300 is built to provide hours of accurate Rubidium holdover after a period
of GNSS discipline. During this holdover period, the GNSS receiver can be
powered off to reduce power consumption and extend battery life. This is done
with HOLDOVER mode. During Holdover, the Rb module maintains frequency
accuracy of ±0.5ppb/day, providing an ideal source for calibrating systems where
a GPS signal is unavailable. After the NR300 has been used as a calibrator, the
unit can be returned to an area where GNSS signal is available, and the GNSS
can be restarted without disciplining the Rubidium. When the GNSS signal is
reacquired, the Rb frequency can be measured, to verify the integrity of the Rb
source.