Status LEDs
There are three status LEDs which provide a quick indication of valid unit
operation.
Alert LED:
The Alert LED will illuminate flashing RED to indicate a GNSS failure
or a power supply failure. The Alert LED will
not
flash RED if any valid input
signal is present.
Oven LED (OCXO option only):
The Oven LED will illuminate RED to indicate
an oven failure.
Fault LED:
The Fault LED will illuminate flashing RED to indicate a channel
fault, when any single channel is outside the user-defined threshold.
Power Supply Status (with optional DC supply)
The power supply status screen provides DC voltage values of the two available
power supply sources. The 90-250V AC input is internally connected to an
internal 24V AC-DC convertor which powers the internal supplies with 24V. This
is diode connected with the external 24V DC input, which can be used in addition
to, or in place of, the AC input.
The values of both DC supplies are measured internally to validate connections.
The measured values of the AC/DC convertor and the DC input are reported in
the third and fourth fields of the second $GPNVS status string.
For details on the status strings, see
Programmer’s Guide.