MODEL DC80T OPERATING INSTRUCTIONS 

In thermocouple temperature simulation the calibrator generates the voltage as an 

actual thermocouple, considering both the simulated temperature (hot joint) and 

connection temperature (cold joint). To ensure stated accuracy the user must use 

a connection cable between calibrator and calibrated device with the same 

thermoelectric properties of the selected thermocouple (extension or 

compensation cable). 
A K type extension cable (yellow connector) is supplied with the calibrator and 

should only be used for K type calibration for maximum accuracy. 
Make sure that the appropriate cable with the correct polarity is being used to 

avoid large calibration errors. 

Thermocouple or mV measurement 

1.

 

Press the power switch   to turn 

on the calibrator. 

2.

 

Press the INPUT / OUTPUT key   

to select INPUT mode. 

3.

 

Press the TC TYPE selection key 

 to configure the instrument to the 

desired input type 

4.

 

Connect the T/C or mV signal to the 

input connector  . 

5.

 

The measured value is presented in 

 

Thermocouple or mV Simulation 

1.

 

Press the power switch    to 

turn on the calibrator. 

2.

 

Press the INPUT / OUTPUT key 

 to select OUTPUT mode. 

3.

 

Press  the TC TYPE selection 

key 

 to configure the desired 

output signal type. 

4.

 

Press the increment / 

decrement keys   

   

 

to adjust the value to be 

simulated. 

5.

 

The simulated signal is 

delivered at the output 

connector  . 

6.

 

To change output values, press 

the increment / decrement keys 

until the desired value is 

presented in the display. 

Changing to other input types is 

accomplished by the TC TYPE 

selection key  . 

 

 

Содержание DC80L

Страница 1: ...L SPECIFICATIONS 9 5 2 MODEL DC80L FRONT PANEL 9 5 3 PROTECTION 10 5 4 MODEL DC80L OPERATING INSTRUCTIONS 10 6 MAINTENANCE 13 1 INTRODUCTION The Novus Calibrators series DC80 are portable instruments...

Страница 2: ...e instrument for measurement or simulation 3 MODEL DC80T The MODEL DC80T is capable of measuring and generating mV and thermocouple signals The specifications shown on the tables below are valid for t...

Страница 3: ...to 50 C Shock Random 2 g 5 Hz to 500 Hz Safety 1 meter drop test Power requirements 6 x AAA 1 5 V Battery Size 205 mm x 98 mm x 46 mm Weight 472 g with batteries 3 2 MODEL DC80T FRONT PANEL The keypad...

Страница 4: ...large calibration errors Thermocouple or mV measurement 1 Press the power switch to turn on the calibrator 2 Press the INPUT OUTPUT key to select INPUT mode 3 Press the TC TYPE selection key to config...

Страница 5: ...target instrument OHM meter of RTD meter connected to the calibrator RTD measurement and simulation specifications Mode Range Accuracy C Excitation current mA Input 4 W Input 2 W 3 W Output Pt10 385...

Страница 6: ...mm x 98 mm x 46 mm Weight 472 g including batteries 4 2 MODEL DC80R FRONT PANEL The keypad and display functions are shown below 4 wire input terminal not used for output 2 wire input output terminal...

Страница 7: ...ding value RTD Simulation 1 Press the power switch to turn ON the calibrator 2 Press the INPUT OUTPUT key to select OUTPUT mode 3 Press RTD mode key to select the desired RTD type 4 Press the Incremen...

Страница 8: ...2 wire transmitter The following tables show the DC80L specifications functions and ranges based on a one year calibration period at ambient temperatures ranging from 18 C to 28 C Counts means units o...

Страница 9: ...2 g 5 Hz to 500 Hz Safety 1 meter drop test Power requirement 6 x AAA Size 204 mm L x 99 mm W x 46 mm H Weight 460 g including batteries External power supply 9 Vdc 500 mA 5 2 MODEL DC80L FRONT PANEL...

Страница 10: ...calibrator Fuse 2 is probably blown out if the calibrator reads 0 000 in mA input mode even under the presence of a current input signal 5 4 MODEL DC80L OPERATING INSTRUCTIONS DC mV and V measurements...

Страница 11: ...egative test lead black to the target equipment 7 For changing the output value use the increment decrement keys and the V mV mode key until the desired value is displayed DC mA measurement a Using ex...

Страница 12: ...splay 4 Plug the red test lead into the LOOP jack and the black one to the mA jack 5 Connect the test leads such that the current enters the red lead and returns to the power supply through the black...

Страница 13: ...ough the black lead 7 To change the output value press the increment decrement keys and or the mA mA mode key Accessories a pair of Test Leads 6 MAINTENANCE Cleaning Periodically wipe the case with a...

Отзывы: