15 | 

P a g e

 

 

11.1

 

   Preparation 

11.1.1

 

Verify that the plug from the AC mains outlet is connected firmly into the connector of the 

power supply units. 

11.1.2

 

Make sure the power cord is connected firmly. 

11.1.3

 

When transmitted light observation is required, pull out the filter system and make the hole 

in the light path. 

11.1.4

 

Be sure to use immersion oil when using fluorescent OIL objectives.  

11.1.5

 

When it is required to interrupt observation for a short period, use the shield in the assistant 

excitation filter part. 

11.1.6

 

Precautions on the specimen colour fading: 

11.1.6.1

 

The system employs high-intensity excitation light to enable bight 

observation of dark fluorescent specimens. As a result, if high-power objectives are 

used frequently colour fading of the specimen occurs early, degrading the view 

(contrast) of fluorescent images. So it is effective to use the shutter frequently to 

avoid illuminating the specimen for a longer period than required. 

11.1.6.2

 

Colour fading of the specimen can also be delayed using commercially 

available colour fading preventing agent (DABCO, etc). The use of colour fading 

preventing agent is recommended when you perform high-magnification 

observation frequently. 

     

 Note that colour fading prevention agent cannot be used with certain specimens. 

12.

 

 

Switching on Electrical Source of the LED driven epi-florescence unit 

 

 

Set the main switch of the power supply unit to “1” (ON).  The LED diode will be at full 

capacity immediately after ignition. 

 

To avoid shortening the burner life, do not turn the burner off and on repeatedly  

 

The LED Diode is at full temperature and efficiency after 30 seconds and the aluminium 

holder can warm considerably after about 10 minutes, it is advisable to leave the unit off for 

approximately 10 minutes before moving the unit to allow sufficient cooling to occur.

 

 

 

 

 

 

 

 

 

 

 

 

Содержание XSZ-107T

Страница 1: ...e YX 2 attachment USER AND OPERATIONAL MANUAL This manual is for the biological microscope XSZ 107T To ensure the safety and optimum performance of this unit please ensure a thorough familiarization w...

Страница 2: ...ram 10 11 9 Adjustment and Operation for brightfield microscopy 12 9 1 Brightness Adjustment 12 10 Focusing on the specimen 13 10 1 Focus on the specimen 13 10 2 Condenser adjustment 13 10 4 Adjusting...

Страница 3: ...ropping to the ground and being damaged The epi fluorescent attachment is a precision instrument b Please ensure that the entire unit is kept out of direct sunlight high temperatures humid or dusty en...

Страница 4: ...r the units as the lenses will get damaged and shall not be replaced under warranty 3 4 If under certain circumstances certain contaminants end up on the lenses i e fingerprints oil smudges one can cl...

Страница 5: ...5 P a g e 4 Components of the microscope...

Страница 6: ...e completed installation of the components The corresponding figures show the assembly steps Before installation ensure the unit is clean as well as every component do not store unpacked units in a ro...

Страница 7: ...op of the microscope body and tighten the holding screw firmly to hold the unit in place 6 2 Installing the binocular head Insert the binocular viewing head into the epi florescence unit and then moun...

Страница 8: ...into the nosepiece from the left or the right side and then revolve the nosepiece clockwise and place one objective per open socket until all the objectives are installed Clean the objectives regularl...

Страница 9: ...its full charge capacity the red light will extinguish and a green light will appear Once in use if one disconnects from the main AC power the internal battery will hold enough current to illuminate...

Страница 10: ...10 P a g e 8 Adjustment Set Diagram...

Страница 11: ...11 P a g e...

Страница 12: ...achment is set to the 0 setting or in the middle setting 9 1 4 Place the specimen slide on the stage in the centre and use the stage swing arm to secure the slide in place 9 1 5 Turn the vertical and...

Страница 13: ...n objective The condenser and objective are coaxial The condenser has been adjusted and pre set before leaving the factory and the maximum height has been pre set so the end user has no need to adjust...

Страница 14: ...nt collar Turn the tension adjustment collar with your fingers When the collar is turned in the direction of the arrow in the figure displayed tension of the coarse adjustment knob increases Turning t...

Страница 15: ...ast of fluorescent images So it is effective to use the shutter frequently to avoid illuminating the specimen for a longer period than required 11 1 6 2 Colour fading of the specimen can also be delay...

Страница 16: ...Hours operation 13 2 Main specifications of the epi florescence attachment YX 2 Epi Fluorescent Illumination Fluorescent Filter Specifications B Exciting Light Filter System G Exciting Light Filter Sy...

Страница 17: ...e nominal voltage Pollution Degree 2 in accordance with IEC 664 Installation Over voltage Category in accordance with IEC 664 12 1 Fluorescent Filter Combinations excitation applications U Auto fluore...

Отзывы: