Section 10
Specifications
General Specifications
46
Model 2001 Service Manual
Rev. 01
General Specifications
•
Operating Conditions:
50-104
°
F (10-40
°
C), 0-90% relative humidity, non-condensing
•
Transport/Storage Conditions:
-10 to +55°C (14-131°F), 10-95% relative humidity, non-condensing
•
Weight: 7 lbs 5 oz. (3.32 kg)
•
Dimensions:
Height, 3.3 inches (8.38 cm) Width, 9 inches (22.86 cm) Depth, 8 inches (20.32 cm)
•
Power: 100-120/200-240 VAC, 50/60 Hz
•
Fuse Rating:
U.S.A.: 0.5 A, 250 V, Slo-Blo (x2) European: T 250 mA/250 V (x2)
•
Battery:
Type, lead-acid gel-cell. Battery Life, 3 hours.
Note: Excessive alerting reduces battery life. When 15 minutes of battery life remain, the
(low battery) indicator illuminates. When the battery becomes exhausted, the monitor
display shuts down. Connect to AC power to recharge battery. Recharge Time, battery
fully recharged in 12-15 hours maximum.
Additional Features
•
2 Minute Silence:
When
AUDIO
key is pressed, deactivates audible alerts for two minutes.
Indicated by illuminated
(2 Min LED).
•
Audio Off:
Feature user selectable. If enabled, press and hold
AUDIO
key for 3 seconds, and
audible alarms will not activate.
Indicated by flashing
(OFF LED).
•
Battery Backed Trend Memory:
Trend memory print of any 30 minutes, 2 hours, 8 hours 12 hours or 24 hours when used
with the Seiko DPU-414 Thermal Printer.
•
Analog (Recorder) Output Module - Optional:
Provides analog output for strip chart applications at the following levels:
Oxygen Saturation value, 10mV/% (100% = 1 V)
Pulse Rate value, 4mV/bpm (250 bpm = 1 V)
Plethysmograph pulse waveform, 0-1V max (AGC)
•
Serial (RS232) Data Output:
Provides RS232 data interface compatible with;
Seiko DPU-414 Thermal Printer
Novametrix Model 1260 Capnograph
Novametrix Model 1010 Telemetry Central Station
RS232 computer interface
NovaCARD
Memory Module
NOVACOM1 Interface
•
Internal Real Time Clock
•
Alert Bar
Содержание MARS PO2 TECH 2001
Страница 2: ......
Страница 4: ...Manufacturing Quality and Safety iv Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 10: ...Section 1 Patient Safety Cautions 4 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 40: ...Section 7 Electronic Tests Safety Testing 34 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 66: ...Section 12 Parts Lists 60 Model 2001 Service Manual Rev 01...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......