Chapter 3: Tuning and Testing Procedures
© 2019 Cutting Edge Optronics
28
Patara II and Iklwa II Service Manual
Figure 2-18: Tightening Hard Plumbing
10.
Connect the coaxial BNC RF cable to the RF driver.
11.
Secure coaxial cable and the humidity sensor cable to the hard plumbing lines.
Figure 2-19: Cables Secured Out of Beam Path
Remove Cover
NOTE:
Gloves must be worn while servicing the laser. Service must be performed in a
clean room environment.
1.
Blow debris off outside of laser housing and cover using nitrogen or Class 0
dry air.
2.
Wipe the housing clean using TechniCloth™ wipes and isopropyl alcohol.
3.
Remove the fasteners and lift the cover from the laser housing.
4.
Do not remove the o-ring from the top groove of the laser housing.