![Northern Meditec Crius V6 Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/northern-meditec/crius-v6/crius-v6_operators-manual_1708588091.webp)
91
/
182
To avoid cross-infection, do not reuse disposable sampling tube.
The used disposable sampling tube should be disposed according to local medical waste
disposal regulations.
Do not through Nomoline sampling tube apply negative pressure (such as using a
syringe) to remove condensate.
Do not grab sampling tube to lift the ISA module or system components, or it may make
off with the ISA module or system components, resulting in ISA module or system
components fall on the patient.
To avoid the ISA module fall on the patient, please ensure it has been fixed when using.
After ISA adult / child model sampling tube access to circuit will increase the patient’s
dead space, and therefore prohibited for infants.
Do not use ISA infant sampling tube for adults, otherwise it will lead to excessive flow
resistance.
If the collection of gas sample need to provide gas for breathing, always use bacteria
filter in the exhaust side.
Ensure ISA module used in electromagnetic environment that this specification defined,
in case of to be affected by mobile or portable communication devices.
ISA monitor or system uses high-frequency electro-surgical devices in the vicinity may
cause interference, and led to incorrect measurements.
ISA monitor banned for MRI environment.
To avoid bacterial filter clogging, do not use ISA monitor and quantitative spray or spray
together.
Check if the sample gas flow rate for given patient type is too high.
If the screen prompts Nomoline blockage information or the input port of ISA sampling
tube is beginning to show red flash, replace the sampling tube.
Do not make any changes to the ISA module, If it has been changed, please do
appropriate tests and inspections to ensure long-term safe operation.
Do not use ISA module external natural cooling function.
If the patient circuit’s positive or negative pressure is too large,which may affect the
sample flow.
If the discharge suction pressure is too large, it may affect the sample flow.
Exhaust gases should be discharged into the drainage system.
ISA module is only designed as an auxiliary method to patient assessment. Please use it
together with other vital signs and symptoms assessed equipment.
Do not stretch ISA monitor probe cable.
To avoid the damage to ISA monitor, should be ensure it has been fixed installed.
7.2.3 Set CO2
In standby state,select
【
Menu
】
->
【
Factory setting
】
->enter the required password->
【
Module Type
Setup
】
.
7.3 Mainstream CO2 module
7.3.1 Installation
1.
Make sure the ventilator is turned off, connect the communication port of IRMA monitor
Содержание Crius V6
Страница 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Страница 15: ...15 182 FOR YOUR NOTES...
Страница 26: ...26 182 FOR YOUR NOTES...
Страница 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Страница 32: ...32 182 FOR YOUR NOTES...
Страница 44: ...44 182 FOR YOUR NOTES...
Страница 56: ...56 182 FOR YOUR NOTES...
Страница 88: ...88 182 FOR YOUR NOTES...
Страница 100: ...100 182 FOR YOUR NOTES...
Страница 125: ...125 182 FOR YOUR NOTES...
Страница 149: ...149 182 FOR YOUR NOTES...
Страница 156: ...156 182 FOR YOUR NOTES...