Nebulization or humidification can increase the resistance of breathing system filters, and
that you need to monitor the filter frequently for increased resistance and blockage.
The ventilation accuracy can be affected by the gas added by use of a nebulizer.
The ventilator shall not be used with nitric oxide.
For non-invasive ventilation, the exhaled volume of the patient can differ from the
measured exhaled volume due to leaks around the mask.
Check if the alarm limit settings are appropriate before taking measurement.
When operating the unit with the power supply unit, always connect the unit to an easily
accessible outlet so that it can be unplugged quickly in the event of a malfunction.
No modification of this equipment is allowed.
Failure to have an alternative means of ventilation such as a self-inflating,
manually-powered resuscitator (as specified in ISO 10651-4) with mask can result in
PATIENT death if the VENTILATOR fails.
Stop using the ventilator and contact us immediately when the buzzer sounds.
Under the ambient temperature of 40ºC, the inspiratory pressure of the ventilator
exceeds 60 cmH2O, and the maximum temperature on the surface of breathing mask may
exceed 41ºC but does not exceed 43ºC
1.1.3 Cautions
CAUTION
The ventilator must be inspected and serviced regularly by trained service personnel.
To ensure patient safety, always prepare resuscitator for use.
Always have a special person attend and monitor the operation of the equipment once
the ventilator is connected to the patient.
During the operation of the ventilator, do not disassemble the inspiration safety valve and
expiration valve unless in standby status.
To ensure patient safety, use only parts and accessories specified in this manual.
At the end of its service life, the equipment, as well as its accessories, must be disposed of
in compliance with the guidelines regulating the disposal of such products.
Magnetic and electrical fields are capable of interfering with the proper performance of
the equipment. For this reason, ensure that all external devices operated in the vicinity of
the equipment comply with the relevant EMC requirements. Mobile phone, X-ray
equipment or MRI devices are a possible source of interference as they may emit higher
levels of electromagnetic radiation.
This system operates correctly at the electrical interference levels identified in this
manual. Higher levels can cause nuisance alarms that may stop mechanical ventilation.
Pay attention to false alarms caused by high-intensity electrical fields.
Содержание Crius V6
Страница 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Страница 15: ...15 182 FOR YOUR NOTES...
Страница 26: ...26 182 FOR YOUR NOTES...
Страница 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Страница 32: ...32 182 FOR YOUR NOTES...
Страница 44: ...44 182 FOR YOUR NOTES...
Страница 56: ...56 182 FOR YOUR NOTES...
Страница 88: ...88 182 FOR YOUR NOTES...
Страница 100: ...100 182 FOR YOUR NOTES...
Страница 125: ...125 182 FOR YOUR NOTES...
Страница 149: ...149 182 FOR YOUR NOTES...
Страница 156: ...156 182 FOR YOUR NOTES...