AUTOLITH

   

System  

Quick

 Reference and Information Guide 

 

Getting STARTED 

 

Turn  on  the  power  “

ON

”  switch  after  plugging  the 

AUTOLITH

®

  System  into  a 

proper power outlet.  Attach the foot-pedal to the front panel foot-pedal port and 
place the sterilized 

EHL Probe

 and 

Extender Cable

 onto the sterile field using 

proper sterile technique. 

Connecting the EHL Probe 

 

Connect  the  EHL  probe  to  the  end  of  the  extender  cable  opposite 
the  square  black  ferrite.    Pass  the  remaining  end  of  the  extender 
cable to the O.R. circulator.   

EHL Probe Recognition 

 

The O.R. circulating nurse should then connect the remaining end of the 
extender  cable  to  the  front  panel  “

PROBE

”  port.    Notice  the  “

Connect 

Probe

” message change to indicate the size of the EHL probe that has 

been connected, i.e. 

1.9Fr. (.63mm)

.  The 

AUTOLITH

®

 System will now 

deliver  the  precise  amount  of  energy  specific  to  the  size  of  the 
connected EHL probe.   

The Endoscope and EHL Probe 

 

The doctor should now prepare to fragment the stone.  Careful insertion of the EHL probe into the 
appropriate endoscope working channel is now accomplished and treatment of the stone is ready 
to  start.    Using  direct  visualization,  the  EHL  probe  tip  and  the  stone  are  sighted.      Normal 
physiological 

SALINE

 (premixed bags of 

.9% normal saline

) should be used for irrigation at all 

times. 

Power 

 

Start 

LOW

 and increase the power if needed for 

effective stone fragmentation. 

 

     

The total power range is 10% to 100% 

      Suggested to start: 30%-40% Power 

®   

               

Indicated for:  

   Urinary & Biliary 

 

             

Stones 

 

Align the green lines to connect 

BiPolar Electro-Hydraulic Lithotripsy 