RV Series Automated Dispensing Systems
48
www.nordsonefd.com [email protected]
+1-401-431-7000 Sales and service of Nordson EFD dispensing systems are available worldwide.
Setting Up the System Using the Robot Initial Setup Wizard
The Robot Initial Setup wizard guides you through all the steps required to properly set up the system, including the
calibration and setting of offsets.
Important: The tip will not rotate, either virtually or physically, until the tool centering calibration portion
(Step 1) of the Robot Initial Setup wizard has been performed.
Opening the Robot Initial Setup Wizard and Setting the Angle of Rotation
#
Click
Step
Reference Image
1
>
• Click SYSTEM SETUP > OPEN.
2
• Click ROBOT INITIAL SETUP.
The Robot Initial Setup wizard opens.
NOTE:
If the optional tip detector or tip
aligner was installed, the wizard shows an
image of the applicable device.
• Perform the actions on tabs 1–6 one at a
time. The actions are also provided in this
manual for your reference as needed.
NOTE:
The wizard buttons change to the
color blue when clicked. All the wizard tabs
include the following two buttons:
- RESET THE 4 AXIS: Click to restart the
wizard from Step 1, using the default
values.
- RESET COLOR: Click to return all
selections on the tab to their default
settings.
Robot Initial Setup wizard
showing the optional tip detector
Robot Initial Setup wizard
showing the optional tip aligner
3
• Enter the desired number of degrees for the
Tip Initial Setup Angle.
The green arrow on the screen changes
based on the entered value. Refer to
“Example of the how the green arrow
changes based on the value entered for
Tip Initial Setup Angle” on page 49 for
images.
NOTE:
Ensure that the angle does not
cause the dispensing device to obscure the
camera view.
4
• Continue to “Robot Initial Setup (Step 1
Tab): Setting Up Tip Detection and Tool
Centering Calibration” on page 49.
Setting Up and Calibrating the System (Required) (continued)