Liquidyn V10D Controller
16
www.nordsonefd.com info@nordsonefd.com
800-556-3484 Sales and service of Nordson EFD dispensing systems are available worldwide.
Setup (continued)
About the V10D Controller Parameters
Refer to "View or Change Parameters" on page 15 to adjust parameters using the V10D software.
Parameter Factory Settings
Program
1
2
3
4
Impulse time (Pulse Time)
2 ms
2 ms
2 ms
2 ms
Time between shots
1000 ms
100 ms
20 ms
10 ms
Set point counter
10
10
10
10
Single shot imp. time
2 ms
2 ms
2 ms
2 ms
Description of Parameters
Parameter
Range of Settings
Description
Impulse time (Pulse Time)
0.1 to 10,000 ms
• Corresponds to the electrical actuation time of the
valve.
• Must be adapted to the corresponding valve type (jet-
type or dot-type).
NOTE:
Refer to"Pulse Time Setting for Jet-Type
Valves" on page 17 or "Pulse Time Setting for Dot-
Type Valves" on page 17 as applicable for your
valves.
• When "Continuous" is disabled (not checked), the
valve remains open until the program is stopped.
Time between shots
1 to 1,000 ms
• Sets the length of the pause between the shots at one
dispensing position, for as long as several shots are
applied.
Set point counter
1 to 65,000
• Sets the number of shots. When "Continuous" is
checked, the valve keeps dispensing until the program
stops.
Single shot imp. time
0.1 to 10,000 ms
• Sets how long the air nozzle remains open after the
valve is closed.