27
System Operation
Charging the System
Display and signal processor must both be charged to use system.
1.
Charge the display and signal processor using the provided Nonin
chargers.
2.
Insert charger into respective display and signal processor charging
ports.
During charging the system may reach temperatures of up to 45 degrees
Celsius, which is safe for contact with healthy adult skin for up to 10 minutes.
Battery
For more information, see the “Internal Power” section in "Specifications"
section on page 56.
WARNING: Do not use the signal processor or display while charging.
Charging is an operator function. To ensure patient safety, the system is
not be in contact with the patient during charging.
CAUTION:
Charging signal processor at temperatures below 0 degrees
°C (32 degrees °F) will decrease battery life. Do not charge signal
processor at temperatures below 0 °C (32 °F).
SP-BLE Battery Indication
Battery Level
1 blinking green light
Critical to Low (0-19%)
1 green light lit, 1 blinking green light
Medium (20-59%)
2 green lights lit, 1 blinking green light
High (60-99%)
3 green lights are lit
Full (100%)
CAUTIONS:
• Follow local, state and national governing ordinances and
recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
• Batteries are a fire hazard if damaged. Do not damage,
mishandle, disassemble, service, or replace with non-specified
components.
• Do not charge Li-Ion batteries at a temperature of 0 °C (32 °F) or
less as this may result in significantly reduced battery life.
!
!
Содержание CO-Pilot H500
Страница 1: ...Operator s Manual CO Pilot Model H500 Wireless Handheld Multi Parameter System English...
Страница 18: ...12 Display Screens and Symbols Figure 2 CO Met Measurement Screen CO Met Disabled 1 3 7 2 9 10 13 12 11 19...
Страница 19: ...13 Display Screens and Symbols Pulse Oximetry Figure 3 SpO2 Measurement Screen 2 3 7 10 11 13 12 1 9...
Страница 20: ...14 Display Screens and Symbols Cerebral and Tissue Oximetry Figure 4 rSO2 Measurement Screen 1 5 8 11 12...
Страница 37: ...31 System Operation Figure 17 Data Transfer Successful...
Страница 66: ...60 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 67: ...61 Technical Information THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 68: ......