
8
Displays, Indicators, and Controls
LCD Indicators and Icons
Pulse Volume
This icon indicates that the pulse volume can be reviewed or adjusted
using the Plus (+) and Minus (-) buttons.
Alarm Volume
This icon indicates that the alarm volume can be reviewed or adjusted
using the Plus (+) and Minus (-) buttons.
Recall Settings
Selecting “Yes” for the Recall Settings icon allows users to retrieve the
previous parameter settings.
Patient Security Mode
This icon can be advanced between NO
(default)
and YES to lock and
unlock the Avant 9700’s alarms. Selecting YES activates Patient Security
mode, locking all alarms to prevent accidental activation or adjustment.
The Plus (+) and Minus (-) buttons must be pressed and held for 3
seconds to advance between NO and YES.
Date and Time
This icon indicates that the
Month, Day, Year
,
Hour
, or
Minute
displays can
be reviewed or adjusted using the Plus (+) and Minus (-) buttons.
Display Intensity
This icon indicates that the LED display intensity can be adjusted using
the Plus (+) and Minus (-) buttons.
Backlight
This icon can be advanced between ON
(default)
and MOM (momentary)
to adjust the display LCD’s backlight.
Data Output
This icon can be advanced between
and
to adjust the output mode
of external printing devices:
•
is used for print-on-demand output
•
is used for real-time (once-per-second) output
Trend Waveform Display Selection
This icon can be advanced between modes 1, 2, and 3 to select data
display ranges. See “Trend Waveform View” for more information.
1
SpO2: 70-100
: 30-250