![Nokeval NSnappy Sense LWEU Series Скачать руководство пользователя страница 4](http://html1.mh-extra.com/html/nokeval/nsnappy-sense-lweu-series/nsnappy-sense-lweu-series_user-manual_1699209004.webp)
4 - Sense LWEU Series
Measuring with NSnappy Sense
Push the sensor probe carefully into the product at least 30 mm deep. Do not use excessive
force, do not for example try to push the probe into a frozen or hard product. Be careful also
when penetrating plastic foils or similar surfaces. The sensor probe can be damaged or bent if
you use excessive force on it.
Wait until the reading has sufficiently stabilized. Note that it can take a few seconds before the
readings start updating in the NSnappy service.
You can speed up the measuring process by pushing the probe into the product before the
actual measuring process. This way the temperature starts to stabilize even before the first
readings appear on the screen.
Measuring with NSnappy Sense IR
Point the lens of the IR sensor directly towards the target object. Measuring distance should be
between 5 to 20 cm. The smaller the target is the shorter the measuring distance should be. The
distance should be at most 4 times the smallest dimension of the target object. The surface of
the target object must be matte or rough in texture and it should not gleam or reflect light.
Shiny or transparent objects cannot be accurately measured.
The reading should stabilize in seconds, but note that it can take a few seconds before the
readings first start updating in the NSnappy service. If the result is not what was expected, make
sure that the measuring distance is as short as possible.
Measurement error of couple of
degrees is possible depending on the surface properties of the target object
.