ASSEMBLING METAL TIP UNIT:
1.
Loosen needle nut until reasonably loose.
2.
Slide needle tip base into the lower slot of the needle
nut.
3.
Screw the needle nut tightly back onto the main
threaded mount.
ASSEMBLING SILICONE TIP UNIT:
1.
The silicone tip
MUST
always be sterilised prior to
use.
2.
Remove the metal tip and needle nut.
3.
Push the silicone tip onto the main threaded mount.
Turn clockwise as you push to ensure it is fi rmly in
place.
Packs of 25 silicone tips are available from your
distributor – Order Code: SP1370B.
CLEANING OR STERILISATION:
The following items may be autoclaved:
Ear tips needle; needle nut; the entire lever action hand
piece – however it is important to apply a small drop of
NJ Phillips Lubricant to any moving part post autoclave
as good maintenance. To apply NJ Phillips Lubricant to
moving parts within the handle grip, carefully unscrew
counter clock wise.
It is unnecessary, nor desirable, to dismantle any part of
the lever action unit for any purpose other than to lubricate
moving parts.
LIQUID MEASURE DISPERSMENT SETTINGS:
1.
To set liquid dispersment
adjust the circular nut
located at the bottom of the
handle. Turn the circular
nut clockwise or counter
clockwise to adjust the
(ml) calibrations printed at
the top of the body post. Adults generally require 3-5 (ml)
setting. The dose is set from the bottom of the lever arm.
2.
Prime instrument to ensure all air is removed prior to
use on the patient.
EAR SYRINGE LEVER ACTION
For use in the removal of cerumen along the outer ear canal.
CIRCULAR NUT
DOSE SETTING
BODY POST
EARDRUM
EAR CANAL
CAUTION:
1.
DO NOT insert the ear tip end into the
outer ear canal.
2.
Take due care - rapid jets of water
directed at the tympanic membrane could
be harmful.
3.
This device should only be used by a
health care professional familiar with ear
syringing techniques.
ARTG No: 145575
The rubber tip is designed to be autoclaved
at temperatures of
up to
250 degrees
celsius for
up to
3.5 hours.
NOTE
Make sure the thread is properly engaged
when reconnecting the handle grip.
To
reconnect handle grip, gently pull lever trigger
until the handle grip cover slides up and over
the handle body and screw clock wise until
tight – DO NOT OVER TIGHTEN!
IMPORTANT
!
DO NOT autoclave the hose piece, rather
wash in a good quality Medical Equipment &
Instrument detergent, then hang to drip dry.
NOTE
AUSTRALIA
NSW / ACT / VIC / WA / TAS
ALL STATES MEDICAL SUPPLIES
Contact:
John Davies
35 Ellen Street
Wollongong
NSW 2500
Tel:
1300 520 510
Fax:
1300 509 089
Email:
QLD
MEDICAL AND SURGICAL REQUISITES PTY. LTD.
Contact:
MEDI + Plus - Service Advisor
33 Fulcrum Street
Richlands, Brisbane
QLD 4077
Tel:
(07) 3859 2900
Fax:
(07) 3859 2995
Email:
SA / NT
Atlas McNeil Healthcare
Unit 5
32 – 50 Weaver Street
Edwardstown
SA 5039
Tel:
1300 741 085
Fax:
1300 365 408
Web:
atlasmcneil.com.au
NEW ZEALAND
ESTEEM HEALTHCARE LTD
Contact:
Mr. Brent Mitchell
PO Box 18-040, Glen Innes
or
24B Morrin Road
Tamaki
Auckland NZ
Tel:
0800 427 968
Fax:
0800 337 833
AS OUR POLICY IS ONE OF CONTINUOUS IMPROVEMENT THE MANUFACTURER RESERVES THE RIGHT TO ALTER THESE SPECIFICATIONS AT ANY TIME.
ALL PRODUCTS PRODUCED BY NJ PHILLIPS PTY LIMITED, ARE IDENTIFIED BY A UNIQUE BATCH NUMBER. THIS IDENTIFICATION NUMBER IS AFFIXED
TO THE PRODUCT TO ALLOW TRACEABILITY BY THE MANUFACTURER AND MUST NOT BE REMOVED IF PRODUCT INTEGRITY IS TO BE MAINTAINED.
NJ PHILLIPS PTY LTD
ABN 36 000 082 002
ADDRESS:
LOCKED BAG 8, GOSFORD, NSW 2250 AUSTRALIA
TELEPHONE:
+61 2 4340 2044
FAX:
+61 2 4340 1991
EMAIL:
MADE IN AUSTRALIA
PAS1043 | QL561-R24
NJ PHILLIPS EAR SYRINGE DISTRIBUTION NETWORK