ASSEMBLING METAL TIP UNIT:

1. 

  Loosen needle nut until reasonably loose.

2.

  Slide needle tip base into the lower slot of the needle 

nut.

3.

  Screw the needle nut tightly back onto the main 

threaded mount.

ASSEMBLING SILICONE TIP UNIT:

1.

 

The silicone tip 

MUST

 always be sterilised prior to 

use.

2.

 

Remove the metal tip and needle nut.

3.

 

Push the silicone tip onto the main threaded mount.  
Turn clockwise as you push to ensure it is fi rmly  in 
place.

Packs of 25 silicone tips are available from your 
distributor – Order Code: SP1370B.

CLEANING OR STERILISATION:

The following items may be autoclaved:

Ear tips needle; needle nut; the entire lever action hand 
piece – however it is important to apply a small drop of 
NJ Phillips Lubricant to any moving part post autoclave 
as good maintenance. To apply NJ Phillips Lubricant to 
moving parts within the handle grip, carefully unscrew 
counter clock wise.

It is unnecessary, nor desirable, to dismantle any part of 
the lever action unit for any purpose other than to lubricate 
moving parts.

LIQUID MEASURE DISPERSMENT SETTINGS:

1.

  To set liquid dispersment 

adjust the circular nut 
located at the bottom of the 
handle. Turn the circular 
nut clockwise or counter 
clockwise to adjust the 
(ml) calibrations printed at 

the top of the body post. Adults generally require 3-5 (ml) 
setting. The dose is set from the bottom of the lever arm.

2.

 

Prime instrument to ensure all air is removed prior to 
use on the patient.

EAR SYRINGE LEVER ACTION

For use in the removal of cerumen along the outer ear canal.

CIRCULAR NUT

DOSE SETTING

BODY POST

EARDRUM

EAR CANAL

CAUTION:

1.

 DO NOT insert the ear tip end into the 

outer ear canal.

2.

 Take due care - rapid jets of water 

directed at the tympanic membrane could 
be harmful.     

3.

 This device should only be used by a 

health care professional familiar with ear 
syringing techniques.

 

ARTG No: 145575

The rubber tip is designed to be autoclaved 
at temperatures of 

up to

 250 degrees 

celsius for 

up to

 3.5 hours.

NOTE

Make sure the thread is properly engaged 
when reconnecting the handle grip. 

To 

reconnect handle grip, gently pull lever trigger 
until the handle grip cover slides up and over 
the handle body and screw clock wise until 
tight – DO NOT OVER TIGHTEN!

IMPORTANT

!

DO NOT autoclave the hose piece, rather 
wash in a good quality Medical Equipment & 
Instrument detergent, then hang to drip dry.

NOTE

AUSTRALIA

NSW / ACT / VIC / WA / TAS

ALL STATES MEDICAL SUPPLIES

Contact:

 John Davies

35 Ellen Street

Wollongong 

NSW 2500

 Tel: 

1300 520 510

    Fax:

 1300 509 089 

Email:

 [email protected]

QLD

MEDICAL AND SURGICAL REQUISITES PTY. LTD. 

Contact: 

MEDI + Plus - Service Advisor

33 Fulcrum Street

Richlands, Brisbane 

QLD 4077

Tel:

 (07) 3859 2900

    Fax: 

(07) 3859 2995

Email:

 [email protected]

SA / NT

Atlas McNeil Healthcare

Unit 5

32 – 50 Weaver Street

Edwardstown 

SA 5039

Tel:

 1300 741 085

    Fax: 

1300 365 408

Web: 

atlasmcneil.com.au

NEW ZEALAND

ESTEEM HEALTHCARE LTD

Contact:

 Mr. Brent Mitchell

PO Box 18-040, Glen Innes 

or 

24B Morrin Road

Tamaki 

Auckland NZ

Tel:

 0800 427 968

    Fax: 

0800 337 833

AS OUR POLICY IS ONE OF CONTINUOUS IMPROVEMENT THE MANUFACTURER RESERVES THE RIGHT TO ALTER THESE SPECIFICATIONS AT ANY TIME. 
ALL PRODUCTS PRODUCED BY NJ PHILLIPS PTY LIMITED, ARE IDENTIFIED BY A UNIQUE BATCH NUMBER. THIS IDENTIFICATION NUMBER IS AFFIXED 
TO THE PRODUCT TO ALLOW TRACEABILITY BY THE MANUFACTURER AND MUST NOT BE REMOVED IF PRODUCT INTEGRITY IS TO BE MAINTAINED.

NJ PHILLIPS PTY LTD  

ABN 36 000 082 002

ADDRESS:

 LOCKED BAG 8, GOSFORD, NSW 2250 AUSTRALIA

TELEPHONE:

 +61 2 4340 2044    

FAX: 

+61 2 4340 1991

EMAIL: 

[email protected]    

MADE IN AUSTRALIA

PAS1043  |  QL561-R24

NJ PHILLIPS EAR SYRINGE DISTRIBUTION NETWORK

Отзывы: