NIVUS Full Pipe
page 68
NFP - rev. 04 / 23.05.2018
31.4 CAL Menu “Simulation”
>Simulation<
Simulation of measurement.
After entering the PIN code select between level, velocity and medi-
um temperature by using the arrow keys >up< and >down<. Press-
ing the arrow keys >left< or >right< will increase or decrease the
simulated flow velocity, level or temperature value in steps of 1 cm or
0.1 °C.
Using ENTER enables to enter the desired simulation value directly.
The flow value which has been calculated by means of the simulated
readings will be indicated on the bottom line of the screen. Relays
which might have been set will switch and programmed mA outputs
supply according current values simultaneously.
In position “h” use the >up< and in position “t” use the >down< arrow
key to toggle between the relay status and the analog output signals
on the bottom line of the screen.
1
Simulated height
2
Simulated flow velocity
3
Simulated medium temperature
4
Calculated simulated flow value
5
Programmed relay activated by simulation
6
Analog output-signals (2
nd
display page; choose by pressing >down< repeatedly)
Fig. 31-6
Simulation mode
Содержание NFP
Страница 82: ...Instruction Manual NIVUS Full Pipe page 82 NFP rev 04 23 05 2018 Approvals and certificates...
Страница 83: ...Approvals and certificates NFP rev 04 23 05 2018 page 83...
Страница 84: ...Instruction Manual NIVUS Full Pipe page 84 NFP rev 04 23 05 2018...
Страница 85: ...Approvals and certificates NFP rev 04 23 05 2018 page 85...
Страница 86: ...Instruction Manual NIVUS Full Pipe page 86 NFP rev 04 23 05 2018...
Страница 87: ...Approvals and certificates NFP rev 04 23 05 2018 page 87...