34 / 44
♦
spa580en22p06
Conditions of correct programming of the data pairs
•
The table must always start with L(1)= 0 and r(1)= output value (assigned to 0 level).
•
The L column may not include identical values.
•
The L and R columns can only have monotonically increased values from top to bottom.
•
If the table contains less than 32 data pairs, the L column must end with a level value ”0” in the row following the last relevant data pair.
i
L
(Left column)
LEVEL VALUES MEASURED
R
(Right column)
OUTPUT VALUE
1
0
R(1)
2
L(2)
R(2)
L(i)
R(i)
nn
L(nn)
R(nn)
nn+1
0
100
EView2 setting:
“Device Settings”
→
“OC-Table”
→
“OCT list”
5.3.10
I
NFORMATIONAL PARAMETERS
(
READ ONLY PARAMETERS
)
P60: - - - - Overall operating hours of the unit (h)
P61: - - - - Time elapsed after last switch-on (h)
P62: - - - - Operating hours of the relay (h)
P63: - - - - Number of switching cycles of the relay
P64: - - - - Actual temperature of the transducer (°C / °F)
If the temperature sensor circuit is broken, the device will give a warning via HART to display “Pt Error” and compensate to +20 ºC (+68 °F).
P65: - - - - Maximum temperature of the transducer (°C / °F)
P66: - - - - Minimum temperature of the transducer (°C / °F)
P70: - - - - Number of Echoes / Echo Map
The amplitude and position of these echoes can also be queried.
P71: - - - - Distance of the of Measuring Window (DIST)
P72 - - - - Amplitude of the selected echo [dB] <0
P73:
Position of the selected echo (time)
[
ms
]