Specifications 

Data bit length

 

Start bit, 1 bit + 8 bits + 2 stop bits + no parity 

Baud rate 

19200bps 

Connector shape

 

USB (B type)   

 
Format 

18 



 

OE 

20 

± 



 

20 

OF 



 

OD 

CAN 

SO 

Space 

Code 

Measurement  Space 

SI 

Unit 

CR 

All data 21 

 

CAN

 

:   

Cancel

 

  * 

In memory mode = number of samples / in normal mode = space

 

SO 

 

Specifies enlarged printout with double width. 

± 

+ tighten/- reverse

 

Measurement code

Includes a decimal point. If no decimal point, a space at the end.

 

 

 

1 0    . 0 0 

    10.00 

SI     

Releases the enlarged printout with double width.

 

Unit 

 

In N•m or otherwise, the rest is a series of spaces.

 

 

 

N • m 

 N

m    ( k g f 

 C m 

 kgf

cm ) 

CR

 

Carriage return

 

6-6.  How to output data   

Data is output every time the “CLEAR” key is pressed after measurement is completed. If the auto 
reset is set, data is automatically output. 
When you save the data using the memory function, the data is output with the procedure given below: 

[1]  Press the “MEM/OUTPUT” key, and the last saved memory number and measurement value 

are displayed in turn. 

[2]  Press the “MEM/OUTPUT” key while the data number and measurement value are displayed. 
[3]  After “FA” (First Address) is displayed on the liquid crystal panel, the memory number is 

displayed. Select the first memory number you want to output using the “UP” or “Down” key, 
and then press the “MEM/OUTPUT” key. 

[4]  After “LA” (Last Address) is displayed on the liquid crystal panel, the memory number is 

displayed. Select the last memory number you want to output using the “UP” or “Down” key, 
and then press the “MEM/OUTPUT” key. 

[5]  “-P-“ is displayed on the liquid crystal panel, and the data is output. When output is finished, 

the machine goes back to the measurement mode. 

Note  When you press the “CLEAR” key for 1 sec or more, the data output mode is cleared. 

6-7.  Changing data output speed 

You can change the speed to output the saved data. 

Note  When you output the data in the high speed output setting, it might result in an error depending 

on your PC and peripherals. If an error occurs, use the normal output mode. 

•  Normal output:    [00] is displayed on the liquid crystal panel.   

High speed output …[01] is 

displayed on the liquid crystal panel.   

*  The default setting is [00].   
•  How to change the output speed   

[1]  Turn on the power by pressing the “ON/OFF” key as holding down the “PRG” key while the 

power is OFF. 

[2]  When you let go the “PRG” key, “00” or “01” is displayed on the liquid crystal panel. Press the 

“UP” key, and the speed setting is switched. 

[3]  After selecting the speed, press the “PRG” key, “-5-“ is displayed on the liquid crystal panel. 

Then, the machine goes back to the torque display, and measurement becomes available. 

Содержание delvo DLT1173A

Страница 1: ...LOI CONTROLLATORE DI COPPIA MANUALE DI FUNZIONAMENTO MOMENTMÄTARE BRUKSANVISNING TORQUÍMETRO MANUAL DE FUNCIONAMIENTO 扭矩测量仪 操作手册 Models DLT1173A DLT1673A DLT1973A Manufactured by NITTO KOHKI CO LTD 2 9 4 Nakaikegami Ohta ku Tokyo 146 8555 Japan Tel 03 3755 1111 Fax 03 3753 8791 Original instructions GB D F I S E C ...

Страница 2: ......

Страница 3: ...nging data output speed 12 6 8 Judgment function and peak down setting 13 6 9 Maximum minimum and average values calculation function 14 6 10 Recharging 14 6 11 Resetting of the system 15 7 Examining and calibrating the Torque Checker 15 Meanings of the symbols and signal terms The meanings of the symbols and signal terms indicated at the top of each precaution Note that even an item described in ...

Страница 4: ...or a necklace or other ornament Any such object may get caught in the rotary part 6 Do not handle the cord roughly Do not move the product while holding its cord when it is attached to the product Do not pull the cord to unplug it Do not let the cord come near heat oil or a sharp edge 7 Fasten the product firmly Before measuring anything that may entail a high torque fasten the product firmly That...

Страница 5: ...in an accident or injury Do not use the product on impact loads 16 Do not disassemble or apply a strong impact or vibration to the product This product is a precision device Therefore do not disassemble it If an excessive impact or vibration damages the Torque Checker it not only will fail to display its performance fully but also may cause an accident or injury 17 Recharge the product correctly R...

Страница 6: ... of the components DLT1173A DLT1673A DLT1973A Charge check lamp Memory display Data output key Average calculation key Judgment setting key Charge check lamp Memory display Data output key Average calculation key Judgment setting key ...

Страница 7: ...cedure Mitsutoyo Digimatic sequential output Analog output Approx 1 5V max Approx 1 0V max Approx 1 1V max No zero offset identical waveforms produced in PP PD and TR modes Sensor system Bending system Spring shape 75 14 1 4 t 75 14 2 8 t 75 14 3 8 t Measuring jig Spring cushion system Method of measurement Measurements taken with the bit installed Using NK bit NK35BN 13 19 10 75 TD20457 Environme...

Страница 8: ...30 4mm diameter type torque range 0 015 to 0 25N m Measuring joint DLW4340 4mm diameter type torque range 0 15 to 1 5N m Measuring joint DLW4350 Hex 5 and hex 6 35 types torque range 0 15 to 1 5N m Measuring joint DLW4360 Hex 5 and hex 6 35 types torque range 0 15 to 6 5N m Measuring joint for screw type DLW4430 No 2 screw type torque range 0 15 to 1 5N m Measuring joint for screw type DLW4530 No ...

Страница 9: ...Setting the measurement mode When you press the measuring mode switch key MODE for approx 1 sec the measuring mode displayed on the liquid crystal panel is changed Every time you press the key measuring mode is switched in turn There are the following three modes You can use an appropriate one according to the use purpose 1 PP peak measurement Usually used to measure the torques of electric screwd...

Страница 10: ...led head screw if necessary 3 Reverse the screwdriver to be measured and loosen the spring of the measuring joint 4 Press the CLEAR key and check that 0 is displayed Note If 0 is not displayed perform zero adjustment 5 Turn the screwdriver to be measured in forward direction Turn it until the screwdriver stops automatically 6 Check that the screwdriver has stopped normally Take the torque reading ...

Страница 11: ...s Do not change the number 8 and the combination direction of disc springs in using a soft joint 3 Checking the measuring joints Be sure to grease up the threaded portion of each measuring joint Check the inside thrust bearing regularly for damage Note In the measurement of ungreased threads it may be that correct measurement is not possible for such reasons as accelerated thread wear Before stora...

Страница 12: ...ile the memory number and measurement value are displayed in turn CLR is blinking on the liquid crystal panel Note When you do not perform any operation while CLR is blinking the measurement becomes available No data is deleted 4 Press the CLEAR key again while CLR is blinking and is displayed on the liquid crystal panel and the selected data is deleted 3 Deletion by specifying the range of the me...

Страница 13: ...aved file 2 Move the decompressed file to the desktop or other location where you can access easily 3 Connect the torque checker to the PC with a USB cable Turn on the power of the torque checker 4 The new hardware detection wizard window is displayed on the PC monitor Select No Not connect now and click Next 5 Go to the next page select Install from a list or specified location and click Next 6 G...

Страница 14: ...emory number you want to output using the UP or Down key and then press the MEM OUTPUT key 4 After LA Last Address is displayed on the liquid crystal panel the memory number is displayed Select the last memory number you want to output using the UP or Down key and then press the MEM OUTPUT key 5 P is displayed on the liquid crystal panel and the data is output When output is finished the machine g...

Страница 15: ...id crystal panel and then the lower limit value displayed 4 To set the lower limit value perform the same operation explained in 2 After the lower limit value is set press the PRG SET key Note When the lower limit value is larger than the upper limit value the judgment is not performed 5 Make the setting of the peak down start value PdLO is displayed on the liquid crystal panel the peak down start...

Страница 16: ... at the bottom of the liquid crystal panel and average value AVE at the bottom of the liquid crystal panel are displayed in this order on the liquid crystal panel When you measure new maximum value minimum value and average value the previously saved data become invalid 6 10 Recharging 1 Automatic power off This product will power itself off if it is left for 10 minutes or if the battery level is ...

Страница 17: ...eset because all the contents of the memory shall be deleted 7 Examining and calibrating the Torque Checker It is recommended to have the product examined about once a year To calibrate it follow the procedure described below We are ready to calibrate it for you if you wish If so consult the shop where you bought the product 1 For a calibration and inspection provide a sash weight pulley and faste...

Страница 18: ... Weight of sash weight kg 0 5 1 5 2 5 5 0 Measured torque N m 0 098 0 490 0 981 1 471 Measured torque N m 0 98 2 94 4 90 9 81 Standard value 1 digit 0 5 F S Standard value 1 digit 0 5 F S DLT1973A Pulley diameter 399mm Weight of sash weight kg 2 0 4 0 6 0 7 5 Measured torque N m 3 92 7 85 11 77 14 71 Standard value 0 5 F S ...

Отзывы: