General information

BO-750 is designed to measure %SpO

2

 and pulse rate accurately, quickly 

and simply at a fingertip using high precision optical components and 
sophisticated analogue-digital circuits.

Haemoglobin in blood turns to oxyhaemoglobin in the lungs to carry oxygen 
to body cells and turn itself to deoxyhaemoglobin after transferring oxygen 
to the body cells and returning to the lungs. These oxyhaemoglobin and 
deoxyhaemoglobin are well known to absorb specific wavelength of lights, 
infrared and visible red, respectively. BO-750 is calibrated to display 
functional oxygen saturation analyzed from pulse wave length detected with 
two high-precision light emitting diodes (LED) emitting infrared and visible 
red lights individually from one side and a photo diode receiving these lights 
passing through fingertip at the other side.

This device is for medical staff use.

Содержание Pulsefit BO-750

Страница 1: ...ormation 2 Warning and precaution 3 Are the following items included 5 Inserting battery 6 Making measurement 8 Performing data communication BO 750BT 12 Error displays and troubleshooting 14 Care and...

Страница 2: ...transferring oxygen to the body cells and returning to the lungs These oxyhaemoglobin and deoxyhaemoglobin are well known to absorb specific wavelength of lights infrared and visible red respectively...

Страница 3: ...per positioning of the device fingertip not correctly placed on LED and under photo sensor movement of fingertip during measurement radiocontrast agent methylene blue indocyanine green indigo carmine...

Страница 4: ...750 on injured or wounded fingers batteries have enough power for long term measurement Exhausted batteries may cause unexpected interruption of measurement during operation Judgment such as change o...

Страница 5: ...t to make a mistake with the strap attachment Hole for strap attachment Finger insertion section Pulse oximeter Strap Single AAA alkaline battery for testing Instruction manual The strap is designed t...

Страница 6: ...mark may not appear correctly All indicators will appear and the initial test will be carried out if the battery is inserted correctly Do not start the measurement during the initial test The display...

Страница 7: ...ors are lit does not mean that the battery is due for replacement Dispose used batteries in a proper manner in accordance with the regulations of each municipality Check the expiry date of the battery...

Страница 8: ...asurement is not possible if the pulse wave cannot be detected normally Do not move your fingertips or body during measurement Your pulse will be disturbed by your physical and mental condition when y...

Страница 9: ...nside of the device When measurement starts the pulse wave is detected and the amplitude intensity is displayed Measurement results are displayed about 8 sec after measurement starts Read the measurem...

Страница 10: ...vel Backlight turns orange when the level is less than 90 SpO2 Unit of blood oxygen saturation level See page 12 See page 7 Press the display switching button to switch the pulse rate display and PI v...

Страница 11: ...sec Auto display switching Invert the device to automatically switch the display If the top and bottom of the display cannot be determined read in the direction that SpO2 can be read correctly Memory...

Страница 12: ...ion Performing communication 1 Tap the pulse oximeter on the app screen 2 Wear the pulse oximeter on your finger and start the measurement 3 Once the measurement results appear press and hold the disp...

Страница 13: ...he app http www nissei kk co jp english If the connection is not smooth remove your finger from the main unit to end the measurement once before re connecting Do not remove the battery from the main u...

Страница 14: ...blood oxygen saturation level Conduct the checks to ensure correct measurement one more time before measuring Measure with another finger if the pulse amplitude intensity is small and the signal is we...

Страница 15: ...ect was stuck in the finger insertion section during the initial test Check that there is nothing in the insertion section and start the measurement after the initial test error display disappears Con...

Страница 16: ...ns humidity shock dust and direct sunlight Do not drop or expose the device to strong shocks Use strap to prevent accidental fall Avoid storing BO 750 in a gaseous atmosphere or places where chemicals...

Страница 17: ...f more than 12 5 mm no protection against water Mode of operation Continuous operation Key to symbols Type BF applied part Refer to instruction manual booklet No SpO2 alarms The used electrical and el...

Страница 18: ...ablishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 N A Voltage fluctuations fli...

Страница 19: ...acy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which BO 750 is used ex...

Страница 20: ...misuse damage caused by unauthorised repair or modification or damage caused by natural disaster violent action or war Purchaser shall bear transport or shipping related costs NISSEI is not liable fo...