Nissei DM-3000 Скачать руководство пользователя страница 17

- 17 -

CAUSE

ERROR SYMBOL/SYMPTOM

REMEDY

Charge the battery. Only a few 

measurements can be made 

before the battery runs down.

Charge the battery. Measurement 

cannot be conducted.

Turn off the device or press 

START/STOP BUTTON or 

SET/MEMORY BUTTON to 

activate the device from sleeping 

mode.

Make sure to turn off the device 

by pressing "  " side of MAIN 

POWER SWITCH after use.

This does not indicate a 

problem.

Make sure that the patient stays 

still during measurement.

Make sure that the patient is 

being measured with correct 

posture.

See NOTES ON BLOOD 

PRESSURE.

LOW BATTERY;

Battery is weak.

BATTERY ERROR;

Battery is exhausted.

ENERGY SAVER MODE;

The device was left turned on 

and has not been operated for 

over approximately 3 minutes.

When the preset pressure 

value was not high enough to 

make measurement or when 

noise was detected, the cuff 

is pressurized again to a 

higher level.

Patient moved during 

measurement.

Patient is not being measured 

with correct posture.

Blood pressure readings 

constantly vary with time of 

measurement and physical and 

mental condition.

Measurement is interrupted 

once and the cuff is 

inflated again.

Blood pressure is different 

each time. The reading is 

extremely low (or high).

The indication appearing at all LCD segment test, 

which is performed when the unit is turned on, does 

not indicate low battery or battery error.

FLASHING

STEADY

ERROR DISPLAYS AND TROUBLESHOOTING

Содержание DM-3000

Страница 1: ...purposes other than those described herein It is essential to read and understand the entire manual before use CONTENTS PARTS AND PRODUCT COMPONENTS 3 GENERAL PRODUCT INFORMATION 5 WARNING PRECAUTION...

Страница 2: ...n the cuff pressure is gradually changed The blood pressure is determined according to the pressure values at appearance and disappearance of the Korotkov sounds Intended purpose and Indication for us...

Страница 3: ...ARK DEFLATION MARK PULSE MARK BATTERY MARK POWER MARK PRESET PRESSURE CONTROL AC INDICATOR MODE SELECTOR SET MEMORY BUTTON START STOP BUTTON MAIN POWER SWITCH OFF ON AC ADAPTOR JACK BATTERY COMPARTMEN...

Страница 4: ...cumference of 32 to 42 cm AIR TUBE AIR PLUG ADAPTOR PLUG BATTERY PLUG AIR CONNECTOR AIR CONNECTOR Order No NPDM3000 010 CUFF SHELL NPDM3000 030 BLADDER NPDM3000 040 AIR PLUG B112784 1 Order No NPDM300...

Страница 5: ...ion of blood pressure can be made either automatically by the oscillometric method or manually by the Auscultatory method using a stethoscope Regular and Large Size Cuffs The regular size cuff will ac...

Страница 6: ...ternally powered equipment Expected service life 5 years Key to Symbols Type BF equipment Class II equipment Important read operating instructions The used electrical and electronic products are not h...

Страница 7: ...to interfered blood flow from cuff inflation make sure that AIR TUBE is not kinking or compressed before measurement Otherwise cuff inflation may not be conducted properly and prolonged Make sure that...

Страница 8: ...rwise injury may be resulted Do not make measurements repeatedly to avoid harmful injury due to interfered blood flow from cuff inflation Do not drop or strike the product Make sure not to expose it t...

Страница 9: ...e screw Connect BATTERY PLUG to BATTERY CONNECTOR inside the battery compartment and insert the battery Put back BATTERY COMPARTMENT COVER and close it by inserting and tightening the screw Connect AC...

Страница 10: ...o be charged Avoid charging the battery if it is not completely exhausted Do not forcibly remove the battery or pull out the plugs holding cords and cables Use manufacturers dedicated AC adaptor and b...

Страница 11: ...ze of cuff for your patient Find a chair and a table so that the patient can be comfortably seated with patient s back and arm supported Legs should not be crossed and feet kept flat on the floor The...

Страница 12: ...is rapidly exhausted from the cuff when measurement is completed Blood pressure and pulse rate are displayed Turn off the device by pressing side of MAIN POWER SWITCH The device will go into energy sa...

Страница 13: ...the segment display The deflation rate can be selected from 2 5 4 5 and 6 5 mmHg sec The set deflation rate returns to the default value of 4 5 when the device is turned off Apply the cuff and stetho...

Страница 14: ...e will go into energy saver mode after approximately 3 minutes even if the SWITCH has not been turned off See page 10 Manually pressurizing the cuff with inflation bulb Disconnect AIR PLUG from DM 300...

Страница 15: ...ed first 2 Press M button to recall stored data one by one Figure following No indicates sequence of data the larger the number the older the data However maximum data can that be stored is 7 Deleting...

Страница 16: ...n off the device or press START STOP BUTTON or SET MEMORY BUTTON to activate the device from sleeping mode OVER PRESSURIZATION The cuff was inflated to the maximum pressure because of movement of body...

Страница 17: ...R Battery is exhausted ENERGY SAVER MODE The device was left turned on and has not been operated for over approximately 3 minutes When the preset pressure value was not high enough to make measurement...

Страница 18: ...not operate when PRESET PRESSURE is set to PUMP OFF Turn PRESET PRESSURE CONTROL and select appropriate pressure between 100 and 280 mmHg for automatic inflation AC INDICATOR also indicates whether t...

Страница 19: ...g Drinking Alcohol Smoking Measurement posture Recent Urination or Bowel Movement Room temperature Changes in the surroundings such as movement or noise etc Blood pressure measured at hospital or clin...

Страница 20: ...r There should be no speaking or moving during blood pressure measurement Otherwise correct measurement results cannot be obtained The ambient temperature should be approximately 20 C for blood pressu...

Страница 21: ...t is stored in the environment with ambient temperature above 40 C or below 10 C please leave it for at least 2 hours before taking a measurement Remove the battery if the device is to be stored for a...

Страница 22: ...d holding START STOP BUTTON down Release the button after the initial display all segment display disappears Two s are displayed and DM 3000 is in CHECKING MODE Inflate the cuff with an inflation bulb...

Страница 23: ...equipment and MRI magnetic resonance imaging equipment etc Use of the device adjacent to or stacked with other equipment must be avoided because it could result in improper operation Use of accessorie...

Страница 24: ...damage caused by natural disaster violent action or war Also cuffs including cuff shells bladders tubes and plugs are not covered by this warranty except for defects in manufacture Purchaser shall be...

Отзывы: