Automatic Sphygmomanometer DM-3000

INSTRUCTIONS

<ENGLISH>

Page

This manual is intended to assist the user for the safe and efficient operation of the Automatic 

Sphygmomanometer DM-3000. The product must be used in accordance with the procedures 

contained in this manual and must not be used for purposes other than those described herein. It is 

essential to read and understand the entire manual before use.

CONTENTS

PARTS AND PRODUCT COMPONENTS   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

3

GENERAL PRODUCT INFORMATION   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

5

WARNING, PRECAUTION, and CONTRAINDICATION  .  .  .  .  .  .

7

PREPARATION FOR USE  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

9

TAKING BLOOD PRESSURE   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

11

MEMORY FUNCTION    .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

15

ERROR DISPLAYS AND TROUBLESHOOTING   .  .  .  .  .  .  .  .  .  .  .  .

16

NOTES ON BLOOD PRESSURE  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

19

CARE AND MAINTENANCE  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

21

TECHNICAL DESCRIPTION FOR ELECTROMAGNETIC 

DISTURBANCES   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

23

WARRANTY AND SERVICE    .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

24

Issued on 10. 27. 2021  A113727-1_7

Содержание DM-3000

Страница 1: ...purposes other than those described herein It is essential to read and understand the entire manual before use CONTENTS PARTS AND PRODUCT COMPONENTS 3 GENERAL PRODUCT INFORMATION 5 WARNING PRECAUTION...

Страница 2: ...n the cuff pressure is gradually changed The blood pressure is determined according to the pressure values at appearance and disappearance of the Korotkov sounds Intended purpose and Indication for us...

Страница 3: ...ARK DEFLATION MARK PULSE MARK BATTERY MARK POWER MARK PRESET PRESSURE CONTROL AC INDICATOR MODE SELECTOR SET MEMORY BUTTON START STOP BUTTON MAIN POWER SWITCH OFF ON AC ADAPTOR JACK BATTERY COMPARTMEN...

Страница 4: ...cumference of 32 to 42 cm AIR TUBE AIR PLUG ADAPTOR PLUG BATTERY PLUG AIR CONNECTOR AIR CONNECTOR Order No NPDM3000 010 CUFF SHELL NPDM3000 030 BLADDER NPDM3000 040 AIR PLUG B112784 1 Order No NPDM300...

Страница 5: ...ion of blood pressure can be made either automatically by the oscillometric method or manually by the Auscultatory method using a stethoscope Regular and Large Size Cuffs The regular size cuff will ac...

Страница 6: ...ternally powered equipment Expected service life 5 years Key to Symbols Type BF equipment Class II equipment Important read operating instructions The used electrical and electronic products are not h...

Страница 7: ...to interfered blood flow from cuff inflation make sure that AIR TUBE is not kinking or compressed before measurement Otherwise cuff inflation may not be conducted properly and prolonged Make sure that...

Страница 8: ...rwise injury may be resulted Do not make measurements repeatedly to avoid harmful injury due to interfered blood flow from cuff inflation Do not drop or strike the product Make sure not to expose it t...

Страница 9: ...e screw Connect BATTERY PLUG to BATTERY CONNECTOR inside the battery compartment and insert the battery Put back BATTERY COMPARTMENT COVER and close it by inserting and tightening the screw Connect AC...

Страница 10: ...o be charged Avoid charging the battery if it is not completely exhausted Do not forcibly remove the battery or pull out the plugs holding cords and cables Use manufacturers dedicated AC adaptor and b...

Страница 11: ...ze of cuff for your patient Find a chair and a table so that the patient can be comfortably seated with patient s back and arm supported Legs should not be crossed and feet kept flat on the floor The...

Страница 12: ...is rapidly exhausted from the cuff when measurement is completed Blood pressure and pulse rate are displayed Turn off the device by pressing side of MAIN POWER SWITCH The device will go into energy sa...

Страница 13: ...the segment display The deflation rate can be selected from 2 5 4 5 and 6 5 mmHg sec The set deflation rate returns to the default value of 4 5 when the device is turned off Apply the cuff and stetho...

Страница 14: ...e will go into energy saver mode after approximately 3 minutes even if the SWITCH has not been turned off See page 10 Manually pressurizing the cuff with inflation bulb Disconnect AIR PLUG from DM 300...

Страница 15: ...ed first 2 Press M button to recall stored data one by one Figure following No indicates sequence of data the larger the number the older the data However maximum data can that be stored is 7 Deleting...

Страница 16: ...n off the device or press START STOP BUTTON or SET MEMORY BUTTON to activate the device from sleeping mode OVER PRESSURIZATION The cuff was inflated to the maximum pressure because of movement of body...

Страница 17: ...R Battery is exhausted ENERGY SAVER MODE The device was left turned on and has not been operated for over approximately 3 minutes When the preset pressure value was not high enough to make measurement...

Страница 18: ...not operate when PRESET PRESSURE is set to PUMP OFF Turn PRESET PRESSURE CONTROL and select appropriate pressure between 100 and 280 mmHg for automatic inflation AC INDICATOR also indicates whether t...

Страница 19: ...g Drinking Alcohol Smoking Measurement posture Recent Urination or Bowel Movement Room temperature Changes in the surroundings such as movement or noise etc Blood pressure measured at hospital or clin...

Страница 20: ...r There should be no speaking or moving during blood pressure measurement Otherwise correct measurement results cannot be obtained The ambient temperature should be approximately 20 C for blood pressu...

Страница 21: ...t is stored in the environment with ambient temperature above 40 C or below 10 C please leave it for at least 2 hours before taking a measurement Remove the battery if the device is to be stored for a...

Страница 22: ...d holding START STOP BUTTON down Release the button after the initial display all segment display disappears Two s are displayed and DM 3000 is in CHECKING MODE Inflate the cuff with an inflation bulb...

Страница 23: ...equipment and MRI magnetic resonance imaging equipment etc Use of the device adjacent to or stacked with other equipment must be avoided because it could result in improper operation Use of accessorie...

Страница 24: ...damage caused by natural disaster violent action or war Also cuffs including cuff shells bladders tubes and plugs are not covered by this warranty except for defects in manufacture Purchaser shall be...