Instruction Manual: EMP-3
NIC-TD-0000067-01
Nippon Instruments Corporation
64
Adjust the reference voltage and signal voltage. Make an adjustment after the instrument has
been stabilized (approximately 30 minutes after the instrument is turned on).
(1)
Turn on the instrument.
(2)
Press “MENU” and then “Maintenance”.
(3)
Proceed to the “Ref/Sig Voltage” screen.
6.5.3.1 AUTOMATIC ADJUSTMENT
The voltages are automatically adjusted to 2.5 V. Both reference voltage and signal voltage
are automatically adjusted.
(1)
Press “AUTO ADJUST”.
(2)
An adjustment is automatically made.
Sequence of Adjustment
The adjustment is made in the following
sequence:
(1)
SPAN of reference voltage
(2)
SPAN of signal voltage
(3)
ZERO of reference voltage
(4)
ZERO of signal voltage