Getting
started
XT V 160 X-ray System
XTM0011-F2
41
you want to use is chosen by clicking the arrow next to the
Live image
button and selecting
the required profile from those displayed.
A Supervisor can create standard imaging profiles. To create or change a profile, click
Options
to open the
C.Clear Options
dialog box. Profiles specify the number of seconds to
average an image, and whether image enhancement, noise reduction, smoothing and/or
sharpening and auto-tone are applied. Refer to the Inspect-X
Online Help
or
User Manuals
for more information.
5.4.6
Position the sample for inspection
You can now move your sample into the path of the X-ray beam with the manipulator to
inspect in real time. Control the manipulator using:
•
The joysticks on the inspection system.
•
The
Axes
controls and readouts in the
Position
panel on the
Position and Optimise
tab.
•
Saved manipulator position sets.
•
The manipulator map on some systems.
•
Interactions with the live image window.
If you are at high magnification, you can find out where you are on your sample by
decreasing magnification.
Using the manipulator, locate your region of interest and zoom in to inspect.
5.4.7
Optimise the image
Generally, the highest contrast in an image is obtained by the lowest X-ray energy (lowest
kV) that produces the desired sample penetration:
•
If you are unable to see detail in your sample, increase the X-ray energy (kV).
Содержание XT V 160
Страница 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 22: ......
Страница 26: ......
Страница 32: ......
Страница 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Страница 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Страница 93: ......