Nikon A1R Скачать руководство пользователя страница 2

Booking:

 

1.

 

When switching off always check the calendar. Users might delete their booking during your 
time  slot.  If  that  happens  to  be  in  the  afternoon  the  system  might  stay.  This  is  potentially 
dangerous and waists the lifetime of the lasers. If you delete your booking please make sure to 
warn the previous user!! 

2.

 

When deleting a booking please send an e-mail to the previous/subsequent user and the institute 
mailing list. 

3.

 

IN-VIVO experiments can be only started in the AFTERNOON as the drift caused by cooling 
and heat within the chamber will affect following users as well. 

System Start up:

 

 

Step 1.:

 Switch on the control computer. 

Step 2.:

 Power up the laser board by turning the key on the controller front panel of the 

laser  unit.  This  will  bring  the  405nm  (Blue)  and  561nm  (Red)  lasers  ON.  The  488nm 
(green) and 647nm (far red) lasers will be in standby mode. 

Содержание A1R

Страница 1: ...Fold the lens cleaning paper three times to form a strip 2 Hold it on the two ends on both sides 3 gently wipe with left right motion Never press with your finger directly 4 Repeat previous step until you cannot see oil stain and repeat once after that one final time 7 When finished ALWAYS turn to the 4x objective and lower the stage to a safe distance Sample Holder 1 When using the universal samp...

Страница 2: ...ng a booking please send an e mail to the previous subsequent user and the institute mailing list 3 IN VIVO experiments can be only started in the AFTERNOON as the drift caused by cooling and heat within the chamber will affect following users as well System Start up Step 1 Switch on the control computer Step 2 Power up the laser board by turning the key on the controller front panel of the laser ...

Страница 3: ...er that is situated to the left of the microscope The switch is on the left side roughly in the middle Step 5 Switch on the stage controller on the table and wait 10 seconds Waiting is important as Nikon microscopes are highly customizable and the microscope body checks which peripheries are currently connected Step 6 Switch on the Microscope body with the RED switch on the optical table Step 7 Op...

Страница 4: ... the 60x oil immersion lens has been used NEVER forget to clean with lens cleaning paper 1 Fold the lens cleaning paper three times to form a strip 2 Hold it on the two ends on both sides 3 Gently wipe with left right motion Never press with your finger directly 4 Repeat previous step until you cannot see oil stain and repeat once after that one final time Step 3 Stop NIS elements If you proceed w...

Страница 5: ...e button on the front panel Step 9 Power OFF the 488nm green and 647nm far red lasers The 647nm far red laser can be powered OFF with the key on the remote control The 488nm Argon Ion laser switched to standby mode by switching the discharge flip switch to OFF on the controller on the laser board unit Step 10 Shut down the laser board with the key on its front panel Step 11 Switch off the computer...