Chapter 25. Capturing High-Speed Images (Resonant Scanner)
A1
81
(6) To use Auto Gain, a function that automatically adjusts the detector sensitivity (HV) based on the set rate
of saturated pixels, click the
[AG] button.
“NG” is displayed for channels that failed in Auto Gain and the HV values are returned to the previous
values.
Use the
[Auto Gain setting] button to change the rate of saturated pixels. Set the maximum value
and minimum value of the rate.
Notes:
•
Auto Gain is disabled during scanning.
•
Auto Gain is disabled during 2Ex1Em or 1Ex2Emx2 line sequence.
•
Auto Gain is disabled when line scan is set.
•
Do not make a manual adjustment in the Acquisition window and do not make an adjustment
using the remote controller during Auto Gain.
(7) Apply Line Average as needed.
Average is a function to scan the same image
multiple times and average it to remove
noises.
Select [Line Average/Integrate] - [Average] in
the Scan setting window and select an
average scan frequency in [Count].
Note: Averaging reduces noise, but decreases
the frame rate (number of images
acquirable per second).
(8) Click the
[XY] button to acquire an image.
25.4
Save an image.
Make the image you want to save active and select [File] - [Save As] to save it.
Note: We recommend that the image be saved in nd2 file format. Conditions such as parameters are
also saved.