Chapter 22. Capturing Virtual Filter Images (Spectral Detector)
A1
67
(4) Click the
[Live] button and adjust [Laser] (laser power), [Si HV] (detector sensitivity), and
[Gain] while checking the image.
Note: The Si HV setting is common to all lasers. Make fine adjustments with the laser power setting.
Note: Using the
[Pixel Saturation Indication] button in the LUTs tab during adjustment makes it
easy to adjust the sensitivity.
Note: If the displayed image is dark, click the
[Auto Scale All] button to adjust the contrast of the
channel automatically to make the image clear.
Note: Turning on and off the [Live] button (scan ON/OFF) during adjustment minimizes fading.
Note: If the LUTs tab is not displayed, right-click the gray area of the software and select
[Visualization Control] - [LUTs] from the displayed menu to call it.
(5) To use Auto Gain, a function that automatically adjusts the detector sensitivity (HV) based on the set rate
of saturated pixels, click the
[AG] button.
“NG” is displayed for channels that failed in Auto Gain and the HV values are returned to the previous
values.
Use the
[Auto Gain setting] button to change the rate of saturated pixels. Set the maximum value
and minimum value of the rate.
Notes:
•
Auto Gain is disabled during scanning.
•
Auto Gain is disabled when line scan is set.
•
Do not make a manual adjustment in the Acquisition window and do not make an adjustment
using the remote controller during Auto Gain.
Using the
[Pixel Saturation Indication] button during
adjustment colors the saturation region.
Auto Scale All