Model No. : ADM-A / ADM-D / ADM-E

Thank you for purchasing the Niigata Seiki Durometer.
Please read this manual thoroughly before use for proper operation.

INSTRUCTION MANUAL

PARTS IDENTIFICATION

TEST SAMPLES

SPECIFICATIONS

Hardness test for rubber and plastic.

APPLICATIONS

This product is compliant with the following 

standards with regards to Indenter shape, size, 

dimensions, and spring force.

JIS K 6253

Rubber, vulcanized or thermoplastic - Determination 

  of hardness

ISO 7619(E)

Rubber-Determination of indentation hardness  by

  means of pocket hardness meters

STANDARDS

 This product is to be used as Hardness Measuring Instrument for 

plastic and rubber. Please use only as directed.

 Only use within applicable measuring range for each model. Use 

beyond specified range may cause errors or damage to instrument.

 Never press indenter against hard surface, such as glass or metal, 

with the exception of testing using the Wear Test Block.

 Do not apply force greater than 2-3 times force specified by the 

relevant standards (JIS, ASTM, ISO, DIN, etc.) Excessive force 

may affect accuracy and cause damage to the instrument.

 Do not apply lateral pressure or move the indenter when in contact 

with test sample.

 Instrument is designed for using in vertical orientation with 

downward pressure. Use in horizontal or upward direction will result 

in inaccurate measurements.

 Please Note –  Measurements taken by this instrument will vary 

from those taken using a different measurement standard, such as 

JIS K6301, etc. 

 This is a precision measurement instrument, Do not drop or apply 

excessive force as it may result in inaccuracy.

 Do not use in wet or oily locations.

 Do not disassemble or modify.

 Do not use organic solvents (thinner, benzine, etc.) to clean.

Please insure that Test Samples meet the following conditions.

 Top and bottom surfaces should be smooth and parallel.

 Sample is free from scratches, air bubbles, and contamination.

 Sample meets the size conditions below:

CAUTION

How To Use (Reverse)

DUROMETER

ADM

-

ADM

-

ADM

-

Wear Test Block

Wear Test

Block

Indenter

Pressure Plate

Test Hole

Bezel

Thumb Rest

Indenter

Pressure

Plate

Indenter

Pressure

Plate

Pointer Follower

Follower Knob

Dial

Pointer

Dial Locking

Screw

A20

A90

Model No.
Hardness Range
Sample Material

Sample Thickness
Indicator Accuracy
Indenter Material
Weight
Accessories

ADM-A

approx. 160g

Above A90

ADM-D

±

1 Scale Resolution

HM35(SKH55)

approx. 160g

Below A20

Rubber,

Soft Plastic,

Tire,

Med. Hardness

Elastomer

Hard Rubber,

Plastic,

Ebonite,

High Hardness

Elastomer

Soft Rubber,

Eraser,

Roll Film,

Low Hardness

Elastomer

6.0mm or greater 6.0mm or greater

10.0mm or greater

ADM-E

approx. 170g

Wear Test Block

Distance to edge: 

12mm

 or greater (ADM-A, ADM-D)

Distance to edge: 

15mm

 or greater (ADM-E)

Test Sample Size Conditions

Thickness:  

6mm

 or greater (ADM-A, ADM-D)

Thickness:

10mm

 or greater (ADM-E)

Measurement Point spacing: 

6mm

 or greater

Measurement Points (5x)