Model No. : ADM-A / ADM-D / ADM-E
Thank you for purchasing the Niigata Seiki Durometer.
Please read this manual thoroughly before use for proper operation.
INSTRUCTION MANUAL
PARTS IDENTIFICATION
TEST SAMPLES
SPECIFICATIONS
・
Hardness test for rubber and plastic.
APPLICATIONS
This product is compliant with the following
standards with regards to Indenter shape, size,
dimensions, and spring force.
JIS K 6253
『
Rubber, vulcanized or thermoplastic - Determination
of hardness
』
ISO 7619(E)
『
Rubber-Determination of indentation hardness by
means of pocket hardness meters
』
STANDARDS
・
This product is to be used as Hardness Measuring Instrument for
plastic and rubber. Please use only as directed.
・
Only use within applicable measuring range for each model. Use
beyond specified range may cause errors or damage to instrument.
・
Never press indenter against hard surface, such as glass or metal,
with the exception of testing using the Wear Test Block.
・
Do not apply force greater than 2-3 times force specified by the
relevant standards (JIS, ASTM, ISO, DIN, etc.) Excessive force
may affect accuracy and cause damage to the instrument.
・
Do not apply lateral pressure or move the indenter when in contact
with test sample.
・
Instrument is designed for using in vertical orientation with
downward pressure. Use in horizontal or upward direction will result
in inaccurate measurements.
・
Please Note – Measurements taken by this instrument will vary
from those taken using a different measurement standard, such as
JIS K6301, etc.
・
This is a precision measurement instrument, Do not drop or apply
excessive force as it may result in inaccuracy.
・
Do not use in wet or oily locations.
・
Do not disassemble or modify.
・
Do not use organic solvents (thinner, benzine, etc.) to clean.
Please insure that Test Samples meet the following conditions.
・
Top and bottom surfaces should be smooth and parallel.
・
Sample is free from scratches, air bubbles, and contamination.
・
Sample meets the size conditions below:
CAUTION
How To Use (Reverse)
DUROMETER
ADM
-
D
ADM
-
A
ADM
-
E
Wear Test Block
Wear Test
Block
Indenter
Pressure Plate
Test Hole
Bezel
Thumb Rest
Indenter
Pressure
Plate
Indenter
Pressure
Plate
Pointer Follower
Follower Knob
Dial
Pointer
Dial Locking
Screw
A20
〜
A90
Model No.
Hardness Range
Sample Material
Sample Thickness
Indicator Accuracy
Indenter Material
Weight
Accessories
ADM-A
approx. 160g
Above A90
ADM-D
±
1 Scale Resolution
HM35(SKH55)
approx. 160g
Below A20
Rubber,
Soft Plastic,
Tire,
Med. Hardness
Elastomer
Hard Rubber,
Plastic,
Ebonite,
High Hardness
Elastomer
Soft Rubber,
Eraser,
Roll Film,
Low Hardness
Elastomer
6.0mm or greater 6.0mm or greater
10.0mm or greater
ADM-E
approx. 170g
Wear Test Block
−
Distance to edge:
12mm
or greater (ADM-A, ADM-D)
Distance to edge:
15mm
or greater (ADM-E)
【
Test Sample Size Conditions
】
Thickness:
6mm
or greater (ADM-A, ADM-D)
Thickness:
10mm
or greater (ADM-E)
Measurement Point spacing:
6mm
or greater
Measurement Points (5x)