61
OPERATING PROCEDURE
Standard Program Refraction
2
22
Enter Night mode and refine the spherical power with the red-green test.
1) Press
The SPH +0.50 D lens is automatically
added to fog the vision. The red-green chart
is presented.
When the cylindrical power is 0.00, the
message whether the cross cylinder test is
skipped or not appears. When the skip is
selected, Steps 22 to 24 are skipped and
the red-green test starts in Step 25.
*14
2) Reduce fogging gradually until the
sharpness of the letters on the red and
green sides appears equal.
Letters on the red side are sharper.
Turn
the dial clockwise one increment.
Letters on the green side are sharper.
Turn the dial counterclockwise one increment.
• In this spherical refinement, a circle of least confusion is maintained on the retina for the
next cylindrical axis measurement with the cross cylinder. When the patient cannot see the
red and green sides equally, make the green side sharper slightly.
23
Measure the cylindrical axis.
1) Press
.
The dots chart is presented.
A cross cylinder lens is placed. The cross cylinder lens type is switched between [Auto], [±0.25],
and [±0.50].
When the cylindrical power of the objective value is 0.25, the same cylindrical power measurement
as Step 24 starts before the cylindrical axis measurement.
*14
When the cylindrical power of the objective value is 0, proceeding the program displays the mes-
sage “No Cylinder power. Skip the Cylinder test? Press [YES] or [NO].”
*15
Selecting “YES” skips
Steps 23 and 24 and then proceeds to Step 25.
• When the cylindrical power is 0 and “NO” is selected for the message “No Cylinder power.
Skip the Cylinder test?” or the cylindrical power becomes 0 during measurement, the
cylindrical power can be additionally reconfirmed by changing the axis.
For details, see the note of "4.2.2 Astigmatism test with cross cylinder lens" (page 157)”.
*14.
When the “Intelligent XC” parameter is set to “Yes” (factory setting), the order or measurement type changes depending
on the CYL value (cylindrical power). See "3.19 Parameter Settings" (page 133) for details.
*15.
When the “Intelligent XC” parameter is set to “Yes” (factory setting), the message is displayed.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...