220
21 Point Exams (only for Plus Package)
21 Point Exams (only for Plus Package)
5.2.8
#14A Cross cylinder (monocular)
Setting contents:
Use chart
Near point cross grid
Set the near point chart II B manually.
Auxiliary lens
(R/L)
Fixed cross cylinder lens (
±
0.5)/Occluder
Set power
Subj (F) + initial addition value
*The initial addition value must be set separately. "l Setting for [#14A C.Cyl(Mon)]" (Page 248)
[Procedure example]
1
With the #7 Subjective value entered, press
[#14A C.Cyl(Mon)].
Left eye numeric value field
Operation message
2
Set the near point chart II B manually at the
40cm position.
3
According to the operation message on the
left bottom of the screen, make the sharp-
ness of horizontal and vertical lines appear
equal with the dial.
When the vertical line is sharp, turn the
dial clockwise and when the horizontal line
is sharp, turn the dial counterclockwise.
4
As the initial value is the right eye, press
or left eye value numeric field after measur-
ing the right eye and measure the left eye in the same manner.
The value displayed in the window is the added power.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...