TEST METHOD
Visual Acuity Test
6
When any of the operations below is performed, Add power VA measurement mode is
finished.
࣭
Press
.
࣭
Switches to the other mode (S/C/A etc.)
4.1.4
Visual Acuity Test using pin hole
The VA can be measured using pin hole of the auxiliary lens.
1
Press
.
The pin hole of the auxiliary lens is set.
For the measurement screen, the VA
mode switches to PHVA and enters VA
measurement mode using pin hole.
2
Test the right, left, and binocular VA.
Press
,
or
and turn the dial
to enter each VA.
The VA can be entered with the mask keys
such as
or
.
3
By switching to the other mode, VA mode using the pin hole is finished.
4.1.5
Input method of the correct character and incorrect character
Pressing
during VA mode displays the +/- function keys. For the entered VA value, +2,
+1, -1, -2 can be added with +/-.
With this function, the character number that can be read or the character number that cannot be read
can be input.
ex) When the VA chart is 4 characters per line,
For the VA 20, if 3 out of 4 characters could be read,
20/20 - 1
For the VA 15, if 1 out of 4 characters could be read,
20/20 + 1
ex) When the VA chart is 5 characters per line,
For the VA 20, if 3 out of 5 characters could be read,
20/20 - 2
For the VA 15, if 3 out of 5 characters could not be read,
20/20 + 2
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...