97
LAN COMMUNICATION
:
Data Format
MLM26*RTZ002F
6
Measurement data
XML file
When the File format parameter is set to “NIDEK_V*.**”
Style sheet file
When the File format parameter is set to “NIDEK_V*.**”
6.4.3
Response to data acquisition
There are two device modes. One is to write data to the shared folder and the other is to confirm data
reception.
Data acquisition is checked by the device monitoring that the XML file of measurement data is no lon-
ger present. On the receiving end, after a data file is produced, the existing data is read out and
deleted or renamed.
The device indicates an error if the XML file remains present after 5 seconds of file writing (The dura-
tion of the time error can be set between 1 to 60 seconds in the parameter setting.)
LM ABCDEFGHIJKLMNOP 20110309103057 012345 . xml
Delimiter (underscore)
Header: Two characters fixed
LM
ID: A maximum of 16 characters
Date and time: 14 characters
(YYYYMMDDHHMMSS)
XML extension:
Three characters fixed
MAC address 3
lower-order bytes:
Six characters
NIDEK LM Stylesheet . xsl
Delimiter (underscore)