8 - 7
10. Put the water into the tank up to the position
illustrated in the right figure.
Even when the filtered water is reused, add
required water.
11. Put the lid on the tank.
12. Attach a new stocking filter to the drain pipe connector.
Stocking filters cannot be reused. Use a new one.
13. Set the drain cover as it was (see the figure
of Step 4).
Place the drain cover in the drain pipe connector on the tank.
Place it with the larger diameter side up.
14. Put the tank back into the cabinet.
15. Connect the feedwater hoses and drain pipe.
Press into the feedwater hose until you hear it click.
NOTE
• As for both PUMP 1 and PUMP 2, be sure to connect the corresponding hoses.
8.2.2 Replacing the water (when the optional FP-100 is not equipped)
Replace the processing water and stocking filter in the tank periodically. It is recommended that
the water be replaced for every 100 lenses processed.
NOTE
• When the message “Please Clean Tank & Pump and Replace Filter” appears and the
processing is stopped, replace the processing water and remove the waste accumulated
on the drain pipe connector.
Stocking filter
Guide of water volume
Drain pipe connector