DEVICE SETTINGS AND MAINTENANCE: Printer Paper Replacement
4.3
Printer Paper Replacement
When a red line appears along the edge of the printer paper, it means that the paper is running short.
In such a case, stop using the printer and replace the printer paper with a new roll.
1
Pull up the cover open lever to open the printer cover and remove the remaining
paper.
2
Insert a new printer paper roll.
Load the printer paper as shown in the picture below. If the roll is loaded with the paper upside down,
printing is not possible.
Set printer paper so that its end extends from the cover.
3
Close the printer cover.
Press the printer cover on both sides to close the cover securely.
CAUTION
• Be sure to use only the printer paper (80620-00001) specified by NIDEK.
If printer paper other than those specified is used, the printer head may be damaged due to printing fail-
ure or paper jam.
• Be sure that printer paper is not loaded in a tilted angle and that the core of the roll is properly
placed.
Printer paper may not be fed properly.
• Be sure that the cover is securely closed.
If the cover is insecurely closed, the auto cutter may not operate properly. In addition, when the print
button is pressed, “ERROR” or “OUT OF PAPER” may appear and printing will not occur.
±
±