![Nidek Medical AFC-330 Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/nidek-medical/afc-330/afc-330_operators-manual_1688053081.webp)
65
OPERATING PROCEDURE
:
Image Capture (Normal Photography Mode [Auto shot, No blink])
2
2.3.2
Image Capture without automatic functions
Image capture with Manual
selected with the Auto/Manual toggle button
All the auto functions are disabled.
Because the auto shot is disabled, the eyelid and eyelashes mark
is not displayed.
1
Perform Steps 1 to 4 of “2.3.1 Image Capture with auto shot” (Page 58).
2
Make the (eight) cornea-reflected spots reflected from the cornea appear.
When the eight cornea-reflected spots are detected, the electronic working dot appears.
3
When the electronic working dot appears, perform alignment in the up, down, right, and
left directions using the joystick so that the electronic working dot is adjusted to the cen-
ter of the target mark.
* If the electronic working dot does not appear (with the required pupil diameter detection
mark displayed as
due to failure to detect the cornea-reflected spots), perform align-
ment so that the circle formed by the cornea-reflected spots and the required pupil diame-
ter mark become concentric.
4
Perform the forward and backward alignment using the joystick so that only a single line
of the indicator is displayed.
For the joystick movement amount, see the working distance indicator and the target mark.
* If the electronic working dot does not appear, the working distance indicator does not
appear even if proper alignment is achieved.
In such a case, perform alignment in the forward and backward directions so that the cor-
nea-reflected spots can be seen most clearly.
Move the joystick back to move the image capturing unit away from the pa-
tient’s eye.
The alignment is optimum.
The indicator is displayed in yellow (not in the fundus observation screen).
Move the joystick forward to move the image capturing unit closer to the pa-
tient’s eye.
Too close to the patient’s eye
Too far from the patient’s eye
When the target mark appears like this, the cornea-reflected spots are not de-
tected. The same indication appears for vertical and horizontal misalign-
ment.
When the target mark appears like this, the cornea-reflected spots are not de-
tected.
Содержание AFC-330
Страница 14: ...XII...
Страница 60: ...44 BEFORE USE Labels and Symbols Common See Before Use Page II...
Страница 68: ...52 BEFORE USE Device and Software Setup...
Страница 174: ...158 MAINTENANCE Consumables and Maintenance Parts List When the lamp is replaced with a spare restock the spare...
Страница 180: ...164 SPECIFICATIONS AND ACCESSORIES Configuration...
Страница 184: ...168 EMC ELECTROMAGNETIC COMPATIBILITY...