1.
Power off the main DC power source or disconnect it from the chassis before installing
any RTIs.
2.
Ensure that the chassis is powered off.
The POWER LED should be off. If it is not off, do not proceed until it is off.
3.
Loosen the screws of the upper rear panel of the chassis.
4.
Position the RTI-12308 at the desired slot and insert the securing screws, but do not fully
tighten them.
5.
Insert an SLSC module into the same slot as its corresponding RTI-12308 while firmly
holding the RTI-12308 in place until the RTI-12308 is firmly connected to the module.
6.
Repeat steps 4 and 5 for all required RTIs.
7.
Tighten the screws for all RTI-12308s and the screws of the upper rear panel of the
chassis.
This ensures proper alignment for future connections between modules and RTIs.
8.
Tighten the two module mounting screws on each newly installed module.
Connecting JH1
To connect JH1, complete the following steps.
1.
Prepare a terminal wire for JH1 by stripping insulation and installing a Molex Ultra-Fit
™
crimp terminal following the manufacturer's specifications.
2.
Insert a prepared terminal into an appropriate terminal socket of Molex Ultra-Fit
™
6-
position receptacle housing.
3.
Repeat steps 1 and 2 for all appropriate terminals.
4.
Insert a receptacle housing into JH1 until the plastic retention latch snaps into place.
JH1 is keyed to prevent reverse installation of the receptacle housing.
2
|
ni.com
|
RTI-12308 Getting Started Guide