NI PXIe-8383mc Скачать руководство пользователя страница 8

8

|

ni.com

|

NI PXIe-8383mc Getting Started Guide

There are also onboard LEDs that report more specific link status which can be useful for 
debugging cases when the front panel PWR/LINK LED reports a solid amber status.

Link Good LEDs

—LEDs that indicate a successful front panel cable link (LED FP) and 

backplane link (LED BP) to the onboard PCIe NTB switch.

Link Activity LEDs

—LED that indicates link activity on the backplane link (LED BP).

Refer to Figure 3 for the location of these LEDs.

Installation and Configuration

This section explains how to unpack and install the NI PXIe-8383mc hardware and software.

Unpacking

The NI PXIe-8383mc is shipped in antistatic packages to prevent electrostatic discharge (ESD) 
to the devices. ESD can damage several components on the device.

Caution

Never touch the exposed pins of connectors. Doing so may damage the 

device.

To avoid such damage in handling the device, take the following precautions:

Ground yourself using a grounding strap or by holding a grounded object.

Touch the antistatic package to a metal part of the computer chassis before removing the 
device from the package.

Remove the device

 

from the package and inspect the device

 

for loose components or any sign 

of damage. Notify NI if the device

 

appears damaged in any way. Do not install a damaged device

 

into the computer or into a PXI Express or CompactPCI Express chassis.

Store the device in the antistatic envelope when not in use.

Hardware Installation

The following are general instructions for installing the NI PXIe-8383mc. Consult your 
computer user manual or technical reference manual for further instructions and warnings.

Caution

The protection provided by the MXI-Express Gen II x8 products can be 

impaired if it is used in a manner not described in this document.

Complete the following steps to install the NI PXIe-8383mc in your PXI Express or 
CompactPCI Express chassis.
1.

Power off your PXI Express or CompactPCI Express chassis, but leave it plugged in while 
installing the NI PXIe-8383mc. The power cord grounds the chassis and protects it from 
electrical damage while you install the module.

Содержание PXIe-8383mc

Страница 1: ...Powering Off the PXImc System 12 Software Installation and Configuration 12 Specifications 13 Worldwide Support and Services 16 About This Guide This document describes the features functions and ope...

Страница 2: ...ss PCIe The terms PCI Express and PCIe are interchangeable in this guide PXI Express chassis In this manual whenever a PXI Express chassis is referenced a CompactPCI Express chassis could be used inst...

Страница 3: ...gure 1 illustrates where the PXImc Logic Block is applied to join the Root Complexes of two systems Figure 1 PXImc System Diagram In the example shown the system on the right is designated as a PXImc...

Страница 4: ...em does not contain a PXImc Logic Block a NI PXIe 8383mc can be installed in the Primary System Host to create a Logical PXImc Device out of the sub system Tip Consolidating all NI PXIe 8383mc adapter...

Страница 5: ...3 NI PXIe 8383mc 153100x 02L Block Diagram The PXImc logic block consists of a non transparent bridge and all the associated circuitry necessary to work with clock references from both the Primary Sys...

Страница 6: ...operating the NI PXIe 8383mc in MC mode and using a NI PXIe 8384 for daisy chain purposes SW4 FACT USER Factory User Mode Switch The FACT USER switch shown in Figure 3 selects between two different me...

Страница 7: ...ion Table 1 describes the function of the front panel LEDs FACT USER Switch Factory User SW4 BAR No Size bytes Address Space Prefetchable FACT BAR 5 32M 32 bit Yes USER BAR 1 BAR 2 BAR4 BAR5 128M 64M...

Страница 8: ...grounding strap or by holding a grounded object Touch the antistatic package to a metal part of the computer chassis before removing the device from the package Remove the device from the package and...

Страница 9: ...jector rail Caution Do not raise the injector ejector handle as you insert the NI PXIe 8383mc It will not insert properly unless the handle is in its downward position so that it does not interfere wi...

Страница 10: ...nsparent bridge in the PXImc Logic Block needs time to self configure before the sub system on the other side of the cable link can connect to it Consolidating all NI PXIe 8383mc adapter modules into...

Страница 11: ...are still powered off Allow time for the operating system to begin loading before proceeding to the next step The three items that follow power sequence B in Figure 4 meet these criteria Caution Do n...

Страница 12: ...mproper power on sequence A reboot of the system that controls the NI PXIe 8383mc may be necessary to resolve this Tip In applications where the power sequence must be automated consider using product...

Страница 13: ...XIe 8384 products NI PXIe 8383mc 153100x 02L Note These specifications are typical at 25 C unless otherwise stated and are subject to change without notice Physical Dimensions 10 0 16 0 cm 3 9 6 3 in...

Страница 14: ...in accordance with IEC 60068 2 1 and IEC 60068 2 2 Meets MIL PRF 28800F Class 3 limits Relative humidity range 5 to 95 noncondensing Tested in accordance with IEC 60068 2 56 Shock and Vibration Opera...

Страница 15: ...uipment is intended for use in commercial light industrial and heavy industrial locations In Europe Canada Australia and New Zealand per CISPR 11 Class A equipment is intended for use only in heavy in...

Страница 16: ...ental information refer to the Minimize Our Environmental Impact web page at ni com environment This page contains the environmental regulations and directives with which NI complies as well as other...