NI PXIe-4304/4305 Calibration Procedure

Documentation

Consult the following documents for information about the PXIe-4304/4305, NI-DAQmx, and 
your application software. All documents are available on 

ni.com

, and help files install with 

the software.

NI PXIe-4304/4305 and TB-4304 User Guide and Terminal Block Specifications

NI-DAQmx driver software installation and hardware setup.

NI PXIe-4304/4305 User Manual

PXIe-4304/4305 usage and reference information.

NI PXIe-4304/4305 Specifications

PXIe-4304/4305 specifications and calibration interval.

NI-DAQmx Readme

Operating system and application software support in NI-DAQmx.

NI-DAQmx Help

Information about creating applications that use the NI-DAQmx driver.

LabVIEW Help

LabVIEW programming concepts and reference information about NI-DAQmx VIs and 
functions.

NI-DAQmx C Reference Help

Reference information for NI-DAQmx C functions and NI-DAQmx C properties.

NI-DAQmx .NET Help Support for Visual Studio

Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET properties, key 
concepts, and a C enum to .NET enum mapping table.

Содержание PXIe-4304

Страница 1: ... 10 Specifications 10 World Wide Support and Services 10 Software Calibrating the NI PXIe 4304 4305 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4304 4305 was first available in NI DAQmx 15 1 1 For the list of devices supported by a specific release refer to the NI DAQmx Readme available on the version specific download page or installatio...

Страница 2: ... reference information NI PXIe 4304 4305 Specifications PXIe 4304 4305 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Information about creating applications that use the NI DAQmx driver LabVIEW Help LabVIEW programming concepts and reference information about NI DAQmx VIs and functions NI DAQmx C Reference Help R...

Страница 3: ...able 1 Table 1 Recommended Equipment for PXIe 4304 4305 Verification and Adjustment Equipment Recommended Model Requirements DMM PXI 4071 Use a DMM that has an accuracy of 32 ppm or better when measuring voltages higher than 10 V and an accuracy of 2 1 μV or better when measuring 0 V PXI Express Chassis PXIe 1062Q If this chassis is unavailable use another PXI Express chassis such as the PXIe 1082...

Страница 4: ...I PXIe 4304 4305 and TB 4304 User Guide and Terminal Block Specifications 2 Configure the PXIe 4139 to voltage output mode and enable remote sensing Connect the PXIe 4139 output to the TB 4304 as shown in Figure 2 Table 2 TB 4304 Analog Signal Names Signal Name Signal Description AI Positive input voltage terminal AI Negative input voltage terminal AIGND Analog ground input NI TB 4304 AI3 AI3 AI2 ...

Страница 5: ...d pair wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The PXIe 4304 4305 temperature will be greater than the ambient temperature Maintain an internal device temperature range of Tcal 5 C for the PXI 4071 1 Keep relative humidity below 80 Allow a warm up time of at least 15 minutes to ensure that the PXIe 4304 4305 measurement circuitry is at a stable opera...

Страница 6: ...the sequence of operations and provides the test points required to verify the PXIe 4304 4305 The verification procedures assume that adequate traceable uncertainties are available for the calibration references The PXIe 4304 4305 has 32 independent analog input channels You can verify the accuracy of any or all of the channels depending upon your desired test coverage Complete the following steps...

Страница 7: ...ement Range 1 V range for test points of 0 V 100 V range for all other test points Digital Resolution 7 5 digits Aperture Time 100 ms Autozero On ADC Calibration On Input Impedance 10 MΩ DC Noise Rejection High Order Number of Averages 1 Power Line Frequency Dependent upon the local power line characteristics Table 5 AI Voltage Mode Setup Configuration Value Channel Name Dev1 aix where x refers to...

Страница 8: ...B 4304 as shown in Figure 2 3 Call the DAQmx Initialize External Calibration function with the following parameters Device In Dev1 Password1 NI 4 Call the 4304 4305 instance of the DAQmx Setup SC Express Calibration function with the following parameters calhandle in calhandle output from DAQmx Initialize External Calibration rangeMax Appropriate Range Max starting with the first Test Point value ...

Страница 9: ... for the remaining Test Point values from Table 7 for the corresponding range that was configured in step 4 10 Repeat steps 4 through 9 for the remaining ranges from Table 7 11 Call the 4304 4305 instance of the DAQmx Adjust SC Express Calibration function with the following parameters calhandle in calhandle output from DAQmx Initialize External Calibration action commit EEPROM Update When an adju...

Страница 10: ...he As Left status of the device Note If any test fails Reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to World Wide Support and Services for assistance in returning the device to NI Specifications Refer to the NI PXIe 4304 4305 Specifications document for detailed PXIe 4304 4305 specification information World Wid...

Отзывы: