Chapter 3
Hardware Overview
3-8
ni.com
In below-low-level analog triggering mode, the trigger is generated when
the signal value is less than
lowValue
.
HighValue
is not used.
Figure 3-4.
Below-Low-Level Analog Triggering Mode
In above-high-level analog triggering mode, the trigger is generated when
the signal value is greater than
highValue
.
LowValue
is not used.
Figure 3-5.
Above-High-Level Analog Triggering Mode
In inside-region analog triggering mode, the trigger is generated when the
signal value is between the
lowValue
and the
highValue
.
Figure 3-6.
Inside-Region Analog Triggering Mode
lowValue
Trigger
highValue
Trigger
highValue
Trigger
lowValue