![NF ZM2371 Скачать руководство пользователя страница 218](http://html1.mh-extra.com/html/nf/zm2371/zm2371_instruction-manual_1684900218.webp)
7.1 Error Messages
7-5
ZM2371/ZM2372
7.1.4 Errors in Remote Control
This section describes main errors that occur in the remote control.
Error
No.
Error message
Cause
0
No error
---
-100
Command error
Command is not correct.
(No detailed classification)
-102
Syntax error
Unrecognizable command or data was received.
-104
Data type error
The format of parameter is improper.
-108
Parameter not
allowed
Too many parameters are used, or a parameter is used
in illegal position.
-109
Missing parameter
The number of parameters is deficient.
-110
Command header
error
Command header is not correct.
(No detailed classification)
-113
Undefined header
Undefined command header is used.
-120
Numeric data error
Numeric data is not correct.
(No detailed classification)
-130
Suffix error
Suffix (multiplier, unit) is not correct.
(No detailed classification)
-140
Character data error
Character data is not correct. (No detailed classification)
-144
Character data too
long
Character data is too long ( > 12 characters).
-150
String data error
Character string data is not correct.
(No detailed classification)
-200
Execution error
Command cannot be executed.
(No detailed classification)
-211
Trigger ignored
Trigger was received but could not be executed.
-221
Settings conflict
Command could not be executed due to the restriction
between plural settings.
-222
Data out of range
Data is out of allowable range.
-241
Hardware missing
Command cannot be executed because hardware to be
operated is not installed.
-300
Device-specific error
Command cannot be executed because of device-specific
internal error. (No detailed classification)
-310
System error
Device-specific internal error occurred.
(For instance, loss of memory data)
-330
Self-test failed
An error was found in the self-test.
-350
Queue overflow
The error queue overflows, and it cannot retain new
error.
-363
Input buffer overrun
The input buffer overflows.
-410
Query
INTERRUPTED
Response to query was interrupted by the subsequent
command.
-420
Query
UNTERMINATED
Response was requested before reception of query is
completed.
-430
Query
DEADLOCKED
Processing cannot be continued because of full buffer.
-440
Query
UNTERMINATED
after indefinite
response
Response is not correct because there is another query
after the query that requests arbitrary length response.
Содержание ZM2371
Страница 1: ...NF Corporation LCR METER ZM2371 ZM2372 Instruction Manual...
Страница 2: ......
Страница 3: ...LCR METER ZM2371 ZM2372 Instruction Manual DA00026021 007...
Страница 250: ......
Страница 252: ......
Страница 253: ......
Страница 254: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...
Страница 255: ...BLANK...