4.8 Setting and Operation of Sweep
WF1967/WF1968
4-76
i) To start sweep with trigger
Set the sweep mode [SwpMode] on the second page of the setting screen as single [Single].
Because a trigger signal is necessary, set a trigger source in trigger [Trig] on the same 2nd page. For
trigger setting,
Set the change time from the starting value to the stop value with sweep time [Time] on the same
2nd page.
Sweep once in single sweep every time a trigger is accepted. If a trigger is received during a
sweep, the sweep will restart from the start value.
The appearance of the change is different as shown in the following figure depending on whether it
is one-way or shuttle sweep function.
In case of one-way sweep, it returns to the start value immediately after the sweep ends.
In case of shuttle sweep, it stands by in sweep terminated state after the sweep ends.
Single one-way sweep
Single shuttle sweep
j) To output waveform only when sweep running
Set the sweep mode [SwpMode] on the second page of the setting screen as gated single [Gated].
This is an operation that combines gate oscillation with sweep. Sweeps start with a trigger. Because
a trigger signal is necessary, set a trigger source in trigger [Trig] on the same second page. For
trigger setting,
Start oscillation/stop phase
Set start oscillation/stop phase in the phase [Phase] on the first page of the setting screen.
However, in phase sweep, the start phase setting is the oscillation starting phase, and the stop phase
setting is the oscillation stop phase.
Stop level
When you want to decide the level while oscillation is stopped apart from the phase, set the stop
level [StpLvl] on the second page of the setting screen as On [On], and then set the level with %
value based on the amplitude full scale. When [Off] is selected, the signal level while oscillation is
stopped is decided by the phase set in [Phase] on the first page of the setting screen. For stop level,
Frequency sweep example
Sweep value
Stop value
Start value
Sweep time
Start value
Trigger
Trigger
Sweep value
Stop value
Start value
Sweep time
Sweep time
Frequency sweep example
Содержание WAVE FACTORY WF1967
Страница 1: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations NF Corporation nbn Austria GmbH...
Страница 2: ......
Страница 3: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations DA00048287 002...
Страница 4: ......
Страница 10: ...WF1967 WF1968 vi...
Страница 19: ...WF1967 WF1968 1 1 1 Overview 1 1 Features 1 2 1 2 Operating Principles 1 4...
Страница 50: ...3 3 Cautions on Floating Ground Connection WF1967 WF1968 3 20 MEMO...
Страница 156: ...4 9 Setting and Operation of Burst WF1967 WF1968 4 106 MEMO...
Страница 206: ...7 9 Identifying Memory Space Required for Saving Arbitrary Waveforms WF1967 WF1968 7 14 MEMO...
Страница 221: ...WF1967 WF1968 9 1 9 Synchronizing Multiple Units 9 1 Connection Procedure 9 2 9 2 Performing Synchronization 9 4...
Страница 226: ...9 2 Performing Synchronization WF1967 WF1968 9 6 MEMO...
Страница 262: ...13 6 Product Information Display Information WF1967 WF1968 13 4 MEMO...
Страница 274: ...14 7 Suspected Failure WF1967 WF1968 14 12 MEMO...
Страница 275: ...WF1967 WF1968 15 1 15 Maintenance 15 1 Outline 15 2 15 2 Operation Inspection 15 4 15 3 Performance Test 15 5...
Страница 286: ...15 3 Performance Test WF1967 WF1968 15 12 MEMO...
Страница 287: ...WF1967 WF1968 16 1 16 List of Initial Settings...
Страница 326: ......
Страница 328: ......
Страница 329: ......