4.2 Settings Memory
FRA5022
4-5
4.2.3 Setting Memory Operation
During a measurement, it is possible to press a numerical key (0 to 9) to select the
corresponding setting.
Regarding AC amplitude and DC bias, switching of setting memory will only change the
setting values. The OUTPUT OSCILLATOR terminal output voltage will not change until
the ON AC/DC key is pressed.
The following operations are possible under the sweep menu or spot menu screen.
■
Selecting setting memory
SETTING memory
Selects a setting memory among the 10 possible.
On the menu screen or setup screen, it is possible to modify the selected setting
memory directly.
■
Setting each item
Except for some, the setup items displayed on the sweep menu screen or spot menu
screen are a subset of the setting memory. The setting items displayed in common with
the sweep menu and spot menu are the items shared in the setting memory. Changing
these items on one menu will also change the corresponding items on the other.
■
Protecting setting mermory
SETTING lock
It is possible to lock a setting memory currently displayed. When locked, all the
settings, except those that are free, will be unavailable.
It is not possible to simultaneously lock or free all of the 10 setting memories.
■
Initializing each setting memory
SETTING clear
Reverts the setting memory contents to the initial values. Does not affect other setting
memories.
Initial values
“Table 3-1 Setting items and initial values”
It is possible to revert all the setup items including the setting memories to their initial
values under the SYSTEM menu screen.
Initialize all
“3.5.4 System Menu Screen”
INITIALIZE all
■
Copying setting memory
SETTING copy to
Copies the contents of the currently displayed setting memory to another setting
memory.
If the displayed memory is locked, the copy destination will be unlocked (free).
If the copy destination is locked, copying results in an error.
Содержание FRA5022
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Страница 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Страница 30: ...FRA5022 2 12...
Страница 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Страница 119: ...FRA5022 5 41...
Страница 131: ...FRA5022 7 8...
Страница 141: ......
Страница 143: ......
Страница 144: ......