11.16 Remote sensing
DP060LM/DP120LM/DP180LM
263
11.16 Remote sensing
This switches the voltage used for measurement. When the remote sensing is on, the sensing input
terminal voltage is used. When it is off, the output terminal voltage is used.
By combining with AGC or Autocal, a voltage drop due to wiring to the load can be compensated.
When the remote sensing is on, the output voltage detection point corrected by the AGC or Autocal
function is switched to the sensing input terminal. When AGC or Autocal is off, only the detection
voltage used for measurement display is switched.
Remote sensing ON or FB setting can be selected in the AC-INT, AC-VCA, AC-SYNC, DC-INT, and
DC-VCA modes. For in the AC mode, it can be selected only when the waveform is a sine wave. It is
turned off when the Sequence or Simulation is selected.
However, in case the stand-by state of sequence mode (step 0) effective only for AC-INT, ACDC-INT
and DC-INT and when the waveform is sine wave or DC. Remote sensing cannot be set to either ON
or FB unless either AC voltage or DC voltage setting is 0 V for ACDC-INT. Also, if the Remote
sensing is ON, remote sensing is turned OFF once when you change ACDC mode from AC mode of
DC mode even in stand-by state. Turn ON it again when you need.
When setting remote sensing to FB, the voltage feedback detection point is changed from inside
(output terminal) to outside (sensing input terminal), and AGC or Autocal cannot be set. Remote
sensing can be set to FB only when DIP or RIN is connected and the function of external control is
set to DevCtrl. For external control, see
11.27
.
External control
Remote
sensing
Voltage feedback
detection point
Measurement
voltage, power, power factor
AGC / Autocal
setting
Enable/Disable
ON
Inside
(output terminal)
Use the sensing input terminal voltage
Enable
OFF
Use the output terminal voltage
DevCtrl
*49
ON
Use the sensing input terminal voltage
OFF
Use the output terminal voltage
FB
*50
Outside
(sensing input
terminal)
Use the sensing input terminal voltage Disable
*49
:
It is used to control DIP or RIN. Put DIP or RIN near this product.
*50
:
It is enabled switching ON
⇔
FB or OFF
⇔
FB only when the output is OFF. The cable connecting
the input terminal of DIP or RIN to the output terminal of this product should be as short as
possible (3 m or less) and use twisted or parallel.
Содержание DP060LM
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Страница 4: ......
Страница 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Страница 24: ......
Страница 215: ...DP060LM DP120LM DP180LM 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202...
Страница 224: ......
Страница 225: ...DP060LM DP120LM DP180LM 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 216...
Страница 239: ...DP060LM DP120LM DP180LM 221 8 Peripherals 8 1 Peripherals Line up 222 8 2 Using in Combination with DIP or RIN 223...
Страница 244: ......
Страница 264: ......
Страница 298: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 280 11 35 Outline Dimensional Drawing Figure 11 2 DP060LM Type2L cabinet...
Страница 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...
Страница 307: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...