11.13 Power Unit Energization Setting
DP060LM/DP120LM/DP180LM
259
Synchronization frequency (only SYNC)
All
models
Display range
38.0 Hz to 525.0 Hz
Resolution 0.1
Hz
Accuracy
0.2 Hz
Harmonic current (AC - INT, fundamental wave 50 Hz/60 Hz only, phase current) *44
Single-phase output
DP060LM DP120LM DP180LM
Effective
value
(rms)
Percent
(%)
Measurement
range
Up to 40th order of the fundamental wave
Full scale
80 A / 40 A
100 %
160 A / 80 A
100 %
240 A / 120 A
100 %
Resolution
0.01 A
0.1 %
0.1 A
0.1 %
Accuracy
(At RMS,
reference
value)
Up to 20th
(1 % of rdg
+ 0.8 A / 0.4 A)
(1 % of rdg
+ 1.6 A / 0.8 A)
(2 % of rdg
+ 2.4 A / 1.2 A)
21st to 40th
(1.5 % of rdg
+ 0.8 A / 0.4 A)
(1.5 % of rdg
+ 1.6 A / 0.8 A )
(3 % of rdg
+ 2.4 A / 1.2 A)
Polyphase output
DP060LM DP120LM DP180LM
Effective
value
(rms)
Percent
(%)
Measurement
range
Up to 40th order of the fundamental wave
Full scale
26.67 A / 13.33 A
100 %
53.33 A / 26.67 A
100 %
80 A / 40 A
100 %
Resolution
0.01 A
0.1 %
0.1 A
0.1 %
Accuracy
(At RMS,
reference
value)
Up to 20th
(1 % of rdg
+ 0.3 A / 0.2 A)
(1 % of rdg
+ 0.55 A / 0.3 A)
(2 % of rdg
+ 0.8 A / 0.4 A)
21st to 40th
(1.5 % of rdg
+ 0.3 A / 0.2 A)
(1.5 % of rdg
+ 0.55 A / 0.3 A)
(3 % of rdg
+ 0.8 A / 0.4 A)
*44
:
The measurement does not conform to the IEC or other standard.
11.13 Power Unit Energization Setting
The power consumption can be decreased by decreasing the number of the power units in operation
according to the load capacity. Each power unit can be enabled (energized)/disabled (not energized)
separately. For the polyphase output, match the number of energized power units for each phase.
Single-phase output
Polyphase output
DP060LM DP120LM
DP180LM
DP060LM
DP120LM DP180LM
Maximum output
power per unit
2 kVA
2 kVA
6 kVA
2 kVA
2 kVA
6 kVA
Number of energized
units setting range
1 to 3
1 to 6
1 to 3
1
1 to 2
1
Energized setting *45 Yes
No
Yes
No
*45
:
It can be set only in case that there are multiple units for each phase.
Содержание DP060LM
Страница 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Страница 4: ......
Страница 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Страница 24: ......
Страница 215: ...DP060LM DP120LM DP180LM 197 5 Description of Screen and Menu 5 1 Screen Configuration 198 5 2 Menu Composition 202...
Страница 224: ......
Страница 225: ...DP060LM DP120LM DP180LM 207 6 Remote Control 6 1 Communication Interface 208 6 2 Remote Local State Switching 216...
Страница 239: ...DP060LM DP120LM DP180LM 221 8 Peripherals 8 1 Peripherals Line up 222 8 2 Using in Combination with DIP or RIN 223...
Страница 244: ......
Страница 264: ......
Страница 298: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 280 11 35 Outline Dimensional Drawing Figure 11 2 DP060LM Type2L cabinet...
Страница 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...
Страница 307: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...